Line 3 was replaced by line 3
- * pump the traps (heat out each for 30 min, but pump longer.
+ * pump the traps (heat out each for 30 min, but pump longer)
Line 5 was replaced by line 5
- system ''and'' on the stick (all voulmes might be connected)
+ system ''and'' on the stick (all volumes might be connected)
Lines 9-10 were replaced by line 9
- * make the seal with clean fingers (a thick layer on the stick
- only a small amount onside the aluminium tube.
+ * make the CAF seal (red silicon paste) with clean fingers
Lines 14-17 were replaced by lines 13-26
- * put into LN2 for precooling, put a small amount of He (1 mbar or
- less) into the IVC.
- * if there is no hurry, the stick might be put directly into
- the cryostat.
+ * put the stick into the Variox or magnet VTI.
+ * insert about 50 mbar 4He into the IVC
+ * take the special assembly to connect the pumping
+ line and the condenser line with the stick
+ * pump out the adapter piece
+ * connect a turbo pump to the IVC, start te pump, but do not yet open the IVC valve.
+
+ The following steps can be done automatically in SEA, under "main diltuion control"
+
+ * ("cool") cool down the heat exchanger of the variox or magnet
+ * When the stick is cold (< 4K, ins SEA "readyforpump") pump the IVC by clicking on "sorbpump" and opening
+ the IVC valve, which is already connected to a running turbo pump.
+ This heats up the VTI to 5 K, which might take a while. After 20 min. with sorb temperature set to 40 K,