Diff: OrangeCryostatCooldownProcedure

Search SINQ Wiki:
SINQ LIN

SINQ Wiki
- Main Page
- Search SINQ Wiki
- Sample Environment
- Probenumg. Intern
- Troubleshooting SICS

This Page
- Page Info
- Printer Friendly

Referenced by
HeCryostats

Wiki Info
- Unused pages
- Undefined pages
- RecentChanges
- Page Index
- System Info
- JSPWiki Docu
- SandBox
- OneMinuteWiki
- Create a New Page




JSPWiki v2.0.52


Difference between version 6 and version 5:
Line 24 was replaced by line 24
- !Cool down
+ !Cool Down
Line 26 was replaced by lines 26-45
- There are two possibilities, each has advantages and drawbacks.
+ # Connect helium pump to the pumping port (P). Switch on the pump and wait until
+ the pressure stabilizes (should be less than a few millibars). Make sure that
+ the warm valve (W) is closed, open cold valve (C) a little, the pressure at the
+ pump should not exceed 20 mbar (30 on a maxi cryostat).
+ In SEA, this could be done by switching the needla valce control to automatic
+ and set 20 - 30 for the flow maximum.
+ # Put a setpoint of 10 K on the heat exchanger.
+ # When the heat exchanger is at 10 K and the sample near 10 K, reduce the pressure
+ by closing the cold valve gradually until the heating power the temperature controller
+ needs to keep 10 K has reached a value between 100 and 200 mW.
+ With SEA, the regulation of the needle valve is done automatically, but the value
+ for "flow set" should be optimized if not known form an earlier run.
+ # Put the temperature setpoint to zero. The cryostat should cool to 1.6 K within 10 minutes.
+ If the base temperature can not be reached or the "nv_Set" is oscillating, most probably
+ the value for "flow set" was not determined properly, and must be increased slighly.
+
+ ''Cool Down with warm valve''
+ In principle there is the possibility to cool down with the use of the warm valve,
+ without He pump. As this is not recommended, it mentioned only for information,
+ and that it has to be made sure that the warm valve has to be closed always.
Removed line 28
- *With He pump
Removed lines 30-47
- ** Connect helium pump to the pumping port (P). Switch on the pump and wait until
- the pressure stabilizes (should be less than a few millibars). Make sure that
- the warm valve (W) is closed, open cold valve (C) a little, the pressure at the
- pump should not exceed 40 mbar (20 on the ILL cold valve controller).
- ** Put a setpoint of 10 K on the heat exchanger.
- ** When the heat exchanger is at 10 K and the sample near 10 K, reduce the pressure
- by closing the cold valve gradually until the heating power the temperature controller
- needs to keep 10 K has reached a value between 100 and 200 mW. Wait half an hour
- and check or change the cold valve setting again.
- ** Put the temperature setpoint to zero. The cryostat should cool to 1.6 K within 10 minutes.
-
- *With warm valve
-
- ** Open cold valve (C), open warm valve (W) by about 3 turns. Close the valve (R) going
- to the He recovery.
- ** When the heat exchanger and the sample temperature have stabilized around at 4.2 K,
- close the warm valve and reopen a little bit, just enough to held 4.2 K.
-
Line 54 was replaced by line 54
- * Dry the sample stick.
+ * Dry the sample stick with a hair blower (not a hot gun!)

Back to OrangeCryostatCooldownProcedure, or to the Page History.