Line 24 was replaced by line 24 |
- !Cool down |
+ !Cool Down |
Line 26 was replaced by lines 26-45 |
- There are two possibilities, each has advantages and drawbacks. |
+ # Connect helium pump to the pumping port (P). Switch on the pump and wait until |
+ the pressure stabilizes (should be less than a few millibars). Make sure that |
+ the warm valve (W) is closed, open cold valve (C) a little, the pressure at the |
+ pump should not exceed 20 mbar (30 on a maxi cryostat). |
+ In SEA, this could be done by switching the needla valce control to automatic |
+ and set 20 - 30 for the flow maximum. |
+ # Put a setpoint of 10 K on the heat exchanger. |
+ # When the heat exchanger is at 10 K and the sample near 10 K, reduce the pressure |
+ by closing the cold valve gradually until the heating power the temperature controller |
+ needs to keep 10 K has reached a value between 100 and 200 mW. |
+ With SEA, the regulation of the needle valve is done automatically, but the value |
+ for "flow set" should be optimized if not known form an earlier run. |
+ # Put the temperature setpoint to zero. The cryostat should cool to 1.6 K within 10 minutes. |
+ If the base temperature can not be reached or the "nv_Set" is oscillating, most probably |
+ the value for "flow set" was not determined properly, and must be increased slighly. |
+ |
+ ''Cool Down with warm valve'' |
+ In principle there is the possibility to cool down with the use of the warm valve, |
+ without He pump. As this is not recommended, it mentioned only for information, |
+ and that it has to be made sure that the warm valve has to be closed always. |
Removed line 28 |
- *With He pump |
Removed lines 30-47 |
- ** Connect helium pump to the pumping port (P). Switch on the pump and wait until |
- the pressure stabilizes (should be less than a few millibars). Make sure that |
- the warm valve (W) is closed, open cold valve (C) a little, the pressure at the |
- pump should not exceed 40 mbar (20 on the ILL cold valve controller). |
- ** Put a setpoint of 10 K on the heat exchanger. |
- ** When the heat exchanger is at 10 K and the sample near 10 K, reduce the pressure |
- by closing the cold valve gradually until the heating power the temperature controller |
- needs to keep 10 K has reached a value between 100 and 200 mW. Wait half an hour |
- and check or change the cold valve setting again. |
- ** Put the temperature setpoint to zero. The cryostat should cool to 1.6 K within 10 minutes. |
- |
- *With warm valve |
- |
- ** Open cold valve (C), open warm valve (W) by about 3 turns. Close the valve (R) going |
- to the He recovery. |
- ** When the heat exchanger and the sample temperature have stabilized around at 4.2 K, |
- close the warm valve and reopen a little bit, just enough to held 4.2 K. |
- |
Line 54 was replaced by line 54 |
- * Dry the sample stick. |
+ * Dry the sample stick with a hair blower (not a hot gun!) |