Diff: OrangeCryostatCooldownProcedure

Search SINQ Wiki:
SINQ LIN

SINQ Wiki
- Main Page
- Search SINQ Wiki
- Sample Environment
- Probenumg. Intern
- Troubleshooting SICS

This Page
- Page Info
- Printer Friendly

Referenced by
HeCryostats

Wiki Info
- Unused pages
- Undefined pages
- RecentChanges
- Page Index
- System Info
- JSPWiki Docu
- SandBox
- OneMinuteWiki
- Create a New Page




JSPWiki v2.0.52


Difference between version 22 and version 4:
At line 0 added 4 lines.
+ !!Orange Cryostat Cooldown Procedure
+
+ For some general remarks on cryogenic handling at the Neutron guide hall [see|http://lns00.psi.ch/sinqwiki/Wiki.jsp?page=CryogenicHandling].
+
At line 1 added 1 line.
+ This is normally done by a sample environment technician
Line 16 was replaced by line 21
- * Increase the temperature of the heat exchanger to 100 K.
+ * Increase the temperature of the heat exchanger to 100 K. (The better part of air will not condense at this temperature)
Lines 20-26 were replaced by lines 25-51
- * Pump down and flush again with He gas, then pump down to a pressure between 10 and 50 mbar.
-
- !Cool down
-
- There are two possibilities, each has advantages and drawbacks.
-
- *With He pump
+ * Pump down and flush again with He gas. To remove all contaminants you should repeat this procedure an infinite number of times (at least 2x).
+ * If you cannot reach 0 mbar (or -1 bar) on the mechanical gauges you might be faced with a leak into the sample chamber or with an offset on the gauge.
+ * Pump down to a pressure between 10 and 50 mbar.
+
+ !Cool Down
+
+ # Connect helium pump to the pumping port (P). Switch on the pump and wait until
+ the pressure stabilizes (should be less than a few millibars). Make sure that
+ the warm valve (W) is closed, open cold valve (C) a little, the pressure at the
+ pump should not exceed 20 mbar (30 on a maxi cryostat).
+ In SEA, this could be done by switching the needle valve control to automatic
+ and set 20 - 30 for the flow maximum.
+ # Put a setpoint of 10 K on the heat exchanger.
+ # When the heat exchanger is at 10 K and the sample near 10 K, reduce the pressure
+ by closing the cold valve gradually until the heating power the temperature controller
+ needs to keep 10 K has reached a value between 100 and 200 mW.
+ With SEA, the regulation of the needle valve is done automatically, but the value
+ for "flow set" should be optimized if not known from an earlier run.
+ # Put the temperature setpoint to zero. The cryostat should cool to 1.6 K within 10 minutes.
+ If the base temperature can not be reached or the "nv_Set" is oscillating, most probably
+ the value for "flow set" was not determined properly, and must be increased slighly.
+
+ ''Cool Down with warm valve''
+
+ In principle there is the possibility to cool down with the use of the warm valve,
+ without He pump. As this is not recommended, it is mentioned only for information.
+ Make sure that the warm valve is always closed except for preparation.
Removed lines 28-44
- ** Connect helium pump to the pumping port (P). Switch on the pump and wait until
- the pressure stabilizes (should be less than a few millibars). Make sure that
- the warm valve (W) is closed, open cold valve (C) a little, the pressure at the
- pump should not exceed 40 mbar (20 on the ILL cold valve controller).
- ** Put a setpoint of 10 K on the heat exchanger.
- ** When the heat exchanger is at 10 K and the sample near 10 K, reduce the pressure
- by closing the cold valve gradually until the heating power the temperature controller
- needs to keep 10 K has reached a value between 100 and 200 mW. Wait half an hour
- and check or change the cold valve setting again.
- ** Put the temperature setpoint to zero. The cryostat should cool to 1.6 K within 10 minutes.
-
- *With warm valve
-
- ** Open cold valve (C), open warm valve (W) by about 3 turns. Close the valve (R) going
- to the He recovery.
- ** When the heat exchanger and the sample temperature have stabilized around at 4.2 K,
- close the warm valve and reopen a little bit, just enough to held 4.2 K.
Line 52 was replaced by lines 60-75
- * Dry the sample stick.
+ * Dry the sample stick with a hair blower (not a hot gun!)
+
+ !He Flow Schematics
+
+ [OrangeCryostatCooldownProcedure/NeedleValveControl.png]
+
+ !Automatic Cryostat Control and Monitoring
+
+ All our cryostats are monitored and controlled with [SEA|http://lns00.psi.ch/sinqwiki/Wiki.jsp?page=Sea].\\
+ All instruments are equipped with a temperature controller (Lakeshore 340) for the temperature control \\
+ and a cryostat control unit (CCU) for the control of pressures, levels and valves.\\
+ The CCU will only work with a correctly configured SEA instance. It can read the pressure in the Helium gas pumping line and set the needle valve accordingly.\\
+ It reads the analog output of AMI Helium level meters and serves as a level indicator for our 2 point liquid Nitrogen level meter.\\
+ The Unit is used to energize the electromagnetic valves involved in the automatic Nitrogen/Helium refill.
+
+ [CCU|CCU.png]

Back to OrangeCryostatCooldownProcedure, or to the Page History.