| At line 2 added 2 lines. |
| + For some general remarks on cryogenic handling at the Neutron guide hall [see|http://lns00.psi.ch/sinqwiki/Wiki.jsp?page=CryogenicHandling]. |
| + |
| Line 19 was replaced by line 21 |
| - * Increase the temperature of the heat exchanger to 100 K. |
| + * Increase the temperature of the heat exchanger to 100 K. (The better part of air will not condense at this temperature) |
| Line 23 was replaced by lines 25-27 |
| - * Pump down and flush again with He gas, then pump down to a pressure between 10 and 50 mbar. |
| + * Pump down and flush again with He gas. To remove all contaminants you should repeat this procedure an infinite number of times (at least 2x). |
| + * If you cannot reach 0 mbar (or -1 bar) on the mechanical gauges you might be faced with a leak into the sample chamber or with an offset on the gauge. |
| + * Pump down to a pressure between 10 and 50 mbar. |
| Line 67 was replaced by line 71 |
| - The CCU will only work with a correctly configured SEA instance. It can read the pressure in the VTI pumping line and set the needle valve accordingly.\\ |
| + The CCU will only work with a correctly configured SEA instance. It can read the pressure in the Helium gas pumping line and set the needle valve accordingly.\\ |
| Line 71 was replaced by line 75 |
| - [CCU|CryogenicHandling/CCU.png] |
| + [CCU|CCU.png] |