Diff: OrangeCryostatCooldownProcedure

Search SINQ Wiki:
SINQ LIN

SINQ Wiki
- Main Page
- Search SINQ Wiki
- Sample Environment
- Probenumg. Intern
- Troubleshooting SICS

This Page
- Page Info
- Printer Friendly

Referenced by
HeCryostats

Wiki Info
- Unused pages
- Undefined pages
- RecentChanges
- Page Index
- System Info
- JSPWiki Docu
- SandBox
- OneMinuteWiki
- Create a New Page




JSPWiki v2.0.52


Difference between version 22 and version 12:
At line 2 added 2 lines.
+ For some general remarks on cryogenic handling at the Neutron guide hall [see|http://lns00.psi.ch/sinqwiki/Wiki.jsp?page=CryogenicHandling].
+
Line 19 was replaced by line 21
- * Increase the temperature of the heat exchanger to 100 K.
+ * Increase the temperature of the heat exchanger to 100 K. (The better part of air will not condense at this temperature)
Line 23 was replaced by lines 25-27
- * Pump down and flush again with He gas, then pump down to a pressure between 10 and 50 mbar.
+ * Pump down and flush again with He gas. To remove all contaminants you should repeat this procedure an infinite number of times (at least 2x).
+ * If you cannot reach 0 mbar (or -1 bar) on the mechanical gauges you might be faced with a leak into the sample chamber or with an offset on the gauge.
+ * Pump down to a pressure between 10 and 50 mbar.
Line 31 was replaced by line 35
- In SEA, this could be done by switching the needla valce control to automatic
+ In SEA, this could be done by switching the needle valve control to automatic
At line 60 added 11 lines.
+
+ !Automatic Cryostat Control and Monitoring
+
+ All our cryostats are monitored and controlled with [SEA|http://lns00.psi.ch/sinqwiki/Wiki.jsp?page=Sea].\\
+ All instruments are equipped with a temperature controller (Lakeshore 340) for the temperature control \\
+ and a cryostat control unit (CCU) for the control of pressures, levels and valves.\\
+ The CCU will only work with a correctly configured SEA instance. It can read the pressure in the Helium gas pumping line and set the needle valve accordingly.\\
+ It reads the analog output of AMI Helium level meters and serves as a level indicator for our 2 point liquid Nitrogen level meter.\\
+ The Unit is used to energize the electromagnetic valves involved in the automatic Nitrogen/Helium refill.
+
+ [CCU|CCU.png]

Back to OrangeCryostatCooldownProcedure, or to the Page History.