Difference between version 36 and version 24:
Lines 2-7 were replaced by lines 2-11
- |used in | ||[VARIOX]||[MA15] ||[MA11] ||[MA02]||[MA09]||[MA6]||[MA7]||[MA10]
- |max. sample diameter* (incl. holder) [[mm] |d|40 |14 |14 |21 |40 |21 |32 |40
- |sample space diameter [[mm] | |42.8|17 |17 |23.1 |42.8 |23.4 |35.8 |42.8
- |distance from [DIL2] mixing chamber to beam center [[mm] |l|83+|296 |296 |342+-5 |272 |295+-5 |n.a. |?
- |distance from [DIL3]/[DIL4] mixing chamber to beam center [[mm]|l|61+|296 |296 |342+-5 |250 |295+-5 |350 |?
- |allowed size below beam center [[mm]|b| |32 |32 |60 |24** |65 |59 |32
+
+ See also [Calculator for Sample Mounting in DIL|Print:CalculatorForSampleMountingInDil]
+
+ |used in | ||[VARIOX]\\out of service||[ORI7]||[MA15] ||[MA11] ||[MA02]||[MA6]||[MA7]||[MA10]|[MB11]|[VarioxB]||[MA09]\\offline use only
+ |max. sample diameter* (incl. holder) [[mm] |d|40 |40 |14 |14 |21 |21 |32 |40 |77 |77 |40
+ |sample space diameter [[mm] | |42.8|42.8 |17 |17 |23.1 |23.4 |35.8 |42.8 |85.8 (77.4) |85.8 (77.4) |42.8
+ |distance from [DIL2] mixing chamber to beam center [[mm] |l|83+|83+ |296 |296 |342+-5 |295+-5 |n.a. |? | | |272
+ |distance from [DIL3]/[DIL4] mixing chamber to beam center [[mm]|l|61+|61+ |296 |296 |342+-5 |295+-5 |350 |? | | |250
+ |distance from [DIL5] mixing chamber to beam center [[mm] |l| | | | | | | | | |210+-10 |210+-10
+ |allowed size below beam center [[mm]|b| | |32 |32 |60 |65 |59 |32 |150 |150 |24**
Line 10 was replaced by line 14
- [Markus Zolliker] for designing sample holders larger than the recommended max. sample diameter.
+ [Markus Zolliker] for designing sample holders slightly larger than the recommended max. sample diameter.
Lines 14-15 were replaced by lines 18-19
- sample holder/weak link there is an M6 hole, 6 mm deep. The (old) weak link
- has a length of m = 245 mm.
+ sample holder/weak link there is an M6 hole, 6 mm deep. The weak link
+ has a length of 30 mm.

Back to DilSampleMounting, or to the Page History.