SEA Help
Known Bugs and Problems
The left part of the SEA client ist blank
Quit the client by closing the window and start a new client.
I can not drive the magnetic field (or any other SE object) from SICS
The command is
drive mf <value>
If you get the following message:
Cannot find mf to drive
ERROR: cannot run mf to 6
then SICS has lost the link to SEA. Enter
connect_sea
from a SICS client. If this does not help, it may be that
the proper device is not selected in SEA.
The needle valve control is unstable
- If you are in automatic mode, switch to controlled mode, and
select the flow by hand (a good flow value is selected when the
heater power at 10 K (stable) is about 0.15 W.
- If you are in controlmode, you might need to increase the "int"
value (to 10 or even 100 sec) and reduce the "minimal pulse" value
to 0.05 sec.
- You might also consider to work in fixed mode (after having found
a reasonable needle valve motor position) when not changing
the temperature over a long time. In fixed mode you can also
move the motor directly from the needle valve controller by means
of the arrow keys.