Removing the Mixture

  • close valve 13A, open valves 9 and 3
  • set the mixing chamber heater and the still heater to 20 mW
  • if the system has to be warmed up to room temperature, the VTI set temperature can be set to 10 K, as soon as the dump pressure is beyond 90 % of its original value (630 mbar)
  • wait until the pressure on P1 drops below 0.01 mbar, then close valves 12, 13, 3 and 9.


This particular version was published on 22-Jun-2007 08:57:34 UTC by MarkusZolliker.