Procedure to Change from/to the Omega Motor on the Sample Stick
Make sure that the serial line (RS232) to the motor controller
is connected and that the motor controller is switched on.
In /home/rita2/rita2_sics/rita2.tcl change the following line
set ma15motor 1 (1 for motor on sample stick, 0 for normal omega table)
and then restart sics with the command
monit restart sicsserver
on the unix prompt.
If the motor is not running, it may be necessary to load the motor
parameters with the el734_test program. Type
el734_test -help
for documentation. A parameter file can be found under /home/users/tascom/PSI/mot_4_01.par
Actually the serial cable for the omega motor is connected to channel 4 (port 3004) on the
terminal server lnsts02.psi.ch.