MA10 vertical field magnet

MA10 replaces MA09, as it has a big sample space (50 mm), possibility of asymmetric field, and works up to 10 Tesla without lambda stage.

Restrictions

maximal radial and axial force: 1000N

ZEBRA 10 Tesla L1 >= 1000 mm
FOCUS 10 Tesla
EIGER 10 Tesla goniometer removed
TASP 10 Tesla goniometer removed, distance to Be filter[1]: > 22 cm
DMC 6 Tesla see elog

Neutron Access

Vertical Access

+-10 deg on a 20 mm high sample

Horizontal access

MA10/MA10_opening_angle.png

Field Homogeneity

MA10/ma10_homogeneity.png

More Information on Intranet Page (drawings): LdmWiki:MA10


Attachments:  ma10_homogeneity.png  MA10_opening_angle.png  taspFilterMA10.png 