| Magnet
 | 
|---|
| field at sample | 9.0 Tesla | 
| " without lambda stage | 7.9 Tesla | 
| operating current for 9 T | < 120 A | 
| maximum allowed vertical magnetic forces | 1000 N (not specified by OI) | 
| maximum allowed horizontal magnetic forces | 500 N (not specified by OI) | 
| homogeneity over a 10 mm diameter sphere | ? | 
| field stability in persistent mode | ? | 
| Stray Field at 9 Tesla
 | 
|---|
| at 1 m horizontal distance | 235 Gauss | 
| at 1 m vertical distance | 310 Gauss | 
| critical distance for loose ferromagnetic parts 500 Gauss/metre
 | 1.15 m vertical 1.10 m horizontal
 | 
| Variable Temperature Insert
 | 
|---|
| temperature range (no field dependend limitations) | 1.8 ... 300 K | 
| temperature stability | +- 0.1 K | 
| needle valve control | manual | 
| Dimensions
 | 
|---|
| sample space diameter | 49.5 | 
| max. useable sample height | 30 mm | 
| split access angle | +- 5 deg | 
| sample height for +-5 deg access(visible by neutrons) | ? | 
| horizontal access angle | > 330 deg | 
| distance from base of cryostat to scattering plane
 | 190 mm | 
| entrance window size (radiussed corners 3 mm) | ? x ? mm | 
| diameter for transfer tube | 9.5 mm | 
| Cryostat
 | 
|---|
| total weight including cryogens | ? kg | 
| useable LHe capacity | ? l | 
| LHe boil off (static mode) | ? cc/hr | 
| LN2 capacity | ? l | 
| typical LN2 boil off | ? cc/hr |