FI

Search SINQ Wiki:
SINQ LIN

SINQ Wiki
- Main Page
- Search SINQ Wiki
- Sample Environment
- Probenumg. Intern
- Troubleshooting SICS

This Page
- Page Info
- Printer Friendly

Referenced by
FurnaceList

Wiki Info
- Unused pages
- Undefined pages
- RecentChanges
- Page Index
- System Info
- JSPWiki Docu
- SandBox
- OneMinuteWiki
- Create a New Page




JSPWiki v2.0.52


This is version 14. It is not the current version, and thus it cannot be edited.
[Back to current version]   [Restore this version]


ILL type furnace

This furnace was produced at the ILL, but is now commercially available at AS Scientific Products Ltd. The heater consists of a double tube of noibium foil, and there are up to 8 radiation screens. These screens help to have a small temperature gradient over the sample. The drawback is, that at lower temperatures the cooling is very weak, and therefore the control difficult.

Current Issues

The furnace is back from repair at AS Scientific Products Ltd. The screens are now in a modified design, which allows an easier replacement of the heater element. However, a repair will take a lot a time.

Power table

When heating up the furnace the first time, ramp up slowly, using the appropriate max. power for the temperature range.

temperaturemax. powerpropintramp
> 1500 2000505020
1300 ... 150010001005020
1100 ... 1300 5002005020
900 ... 1100 3162005010
700 ... 900 158100205
500 ... 700 15810025



Attachments:  illofen2.jpg  illofen3.jpg  illofen4.jpg  illofen1.jpg 
This particular version was published on 29-Jan-2009 15:17:40 UTC by MarkusZolliker.