+ !The needle valve control is unstable |
+ * If you are in automatic mode, switch to controlled mode, and |
+ select the flow by hand (a good flow value is selected when the |
+ heater power at 10 K (stable) is about 0.15 W. |
+ * If you are in controlmode, you might need to increase the "int" |
+ value (to 10 or even 100 sec) and reduce the "minimal pulse" value |
+ to 0.05 sec. |
+ * You might also consider to work in fixed mode (after having found |
+ a reasonable needle valve motor position) when not changing |
+ the temperature over a long time. In fixed mode you can also |
+ move the motor directly from the needle valve controller by means |
+ of the arrow keys. |