| At line 24 added 13 lines. |
| + |
| + !The needle valve control is unstable |
| + * If you are in automatic mode, switch to controlled mode, and |
| + select the flow by hand (a good flow value is selected when the |
| + heater power at 10 K (stable) is about 0.15 W. |
| + * If you are in controlmode, you might need to increase the "int" |
| + value (to 10 or even 100 sec) and reduce the "minimal pulse" value |
| + to 0.05 sec. |
| + * You might also consider to work in fixed mode (after having found |
| + a reasonable needle valve motor position) when not changing |
| + the temperature over a long time. In fixed mode you can also |
| + move the motor directly from the needle valve controller by means |
| + of the arrow keys. |