| Line 1 was replaced by line 1 |
| - !!SEA Help |
| + !!!SEA Help |
| Line 3 was replaced by line 3 |
| - !Known Bugs and Problems |
| + !!Known Bugs and Problems |
| Line 5 was replaced by lines 5-6 |
| - ''Problem: The left part of the SEA client ist blank'' |
| + !The left part of the SEA client ist blank |
| + Quit the client by closing the window and start a new client. |
| Lines 7-11 were replaced by lines 8-9 |
| - Help: Quit the client by closing the window and start a new client |
| - |
| - ''Problem: I can not drive the magnetic field from SICS'' |
| - |
| - Help: The command is |
| + !I can not drive the magnetic field (or any other SE object) from SICS |
| + The command is |
| At line 26 added 28 lines. |
| + |
| + !The needle valve control is unstable |
| + * If you are in automatic mode, switch to controlled mode, and |
| + select the flow by hand (a good flow value is selected when the |
| + heater power at 10 K (stable) is about 0.15 W. |
| + * If you are in controlled mode, you might need to increase the "flow int" |
| + value (to 10 or even 100 sec) and change the pulse length. |
| + * You might also consider to work in fixed mode (after having found |
| + a reasonable needle valve motor position) when not changing |
| + the temperature over a long time. |
| + |
| + !Restarting SEA |
| + Restarting the SEA client is sometimes a good idea when problems |
| + arise. Restarting the SEA server is rarely necessary, but might also |
| + help. From the instrument account enter |
| + {{{ |
| + monit restart sea |
| + }}} |
| + to restart the SEA server. Normally, restarting the SEA server does |
| + not harm the instrument program. However, when the SICS server is in |
| + Driving mode, the objects linked to SEA may be lost. In this case |
| + just enter |
| + {{{ |
| + connect_sea |
| + }}} |
| + on the SICS prompt. |
| + |
| + |