At line 2 added 10 lines. |
+ It is highly recommended to use the automatic scripts in SEA. |
+ You might choose between: |
+ |
+ * remove_only: remove the mixture and stay at 10 K |
+ * remove_and_warmup: remove and warm up to 80 K for removing the stick |
+ * remove_and_condense |
+ * remove_and_sorbump |
+ |
+ Manual procedure: |
+ |
Line 4 was replaced by lines 14-15 |
- * set the mixing chamber heater and the still heater to 20 mW |
+ * set the mixing chamber heater and the still heater to 19 mW |
+ * set the VTI temperature to 4 K |
Lines 8-10 were replaced by lines 19-22 |
- * wait until the pressure on P1 drops below 0.01 mbar (or 0.1 mbar, |
- if the stick is used again), then close all valves, and stop the |
- 3He pump. |
+ * wait until the pressure on P1 drops below 0.1 mbar, then close |
+ all valves, and stop the 3He pump. |
+ |
+ [DIL1/igh.png] |