| Line 1 was replaced by line 1 |
| - !Procedure to Change from/to the Omega Motor on the Sample Stick |
| + !!Procedure to Change the Option for the Sample Stick Rotation |
| Lines 3-6 were replaced by lines 3-6 |
| - Make sure that the serial line (RS232) to the motor controller |
| - is connected and that the motor controller is switched on. |
| - The cables have to be connected, before the motor controller is switched on, |
| - otherwise you will not be able to drive the motor. |
| + There are 7 options now: |
| + ;{{set stickmotor 0}}:''Normal mode: motor __om__ is sample table'' |
| + ;{{set stickmotor 1}}:''MA15 mode: motor __om__ is sample stick, motor __sro__ is sample table'' |
| + ;{{set ma15motor 2}}:''MA02 mode: motor __om__ is sample table, motor __stick__ is sample stick'' |
| Line 8 was replaced by lines 8-9 |
| - In /home/rita2/rita2_sics/rita2.tcl change the following line |
| + !Change to Omega on Sample Stick (for MA15) |
| + * In {{/home/rita2/rita2_sics/rita2.tcl}} change the following line |
| Line 10 was replaced by line 11 |
| - set ma15motor 1 (1 for motor on sample stick, 0 for normal omega table) |
| + set stickmotor 15 |
| Line 12 was replaced by line 13 |
| - and then restart sics with the command |
| + * Stop the SICS server at the unix prompt: |
| Line 14 was replaced by line 15 |
| - monit restart sicsserver |
| + monit stop sicsserver |
| Line 16 was replaced by lines 17-26 |
| - on the unix prompt. |
| + * Switch off the motor controller (if not already off) |
| + * Make sure that the cables to the rotation unit are connected, and |
| + that the serial line (RS232) to the motor controller is connected. |
| + * Switch on the motor controller. |
| + * Start sics with the command |
| + {{{ |
| + monit start sicsserver |
| + }}} |
| + |
| + ''Outdated information:'' |
| Line 21 was replaced by line 31 |
| - el734_test -help |
| + el734_test -help |
| Line 23 was replaced by line 33 |
| - for documentation. A parameter file can be found under /home/users/tascom/PSI/mot_4_01.par |
| + for documentation. A parameter file can be found under /home/users/tascom/PSI/mot_4_01.par (no longer there!) |