Line 1 was replaced by line 1 |
- !Procedure to Change from/to the Omega Motor on the Sample Stick |
+ !!Procedure to Change the Option for the Sample Stick Rotation |
Line 3 was replaced by lines 3-21 |
- in /home/rita2/rita2_sics/rita2.tcl change the following line |
+ There are 7 options now: |
+ ;{{set stickmotor 0}}:''Normal mode: motor __om__ is sample table'' |
+ ;{{set stickmotor 1}}:''MA15 mode: motor __om__ is sample stick, motor __sro__ is sample table'' |
+ ;{{set ma15motor 2}}:''MA02 mode: motor __om__ is sample table, motor __stick__ is sample stick'' |
+ |
+ !Change to Omega on Sample Stick (for MA15) |
+ * In {{/home/rita2/rita2_sics/rita2.tcl}} change the following line |
+ {{{ |
+ set stickmotor 15 |
+ }}} |
+ * Stop the SICS server at the unix prompt: |
+ {{{ |
+ monit stop sicsserver |
+ }}} |
+ * Switch off the motor controller (if not already off) |
+ * Make sure that the cables to the rotation unit are connected, and |
+ that the serial line (RS232) to the motor controller is connected. |
+ * Switch on the motor controller. |
+ * Start sics with the command |
Line 5 was replaced by line 23 |
- set ma15motor 1 (1 for motor on sample stick, 0 for normal omega table) |
+ monit start sicsserver |
Line 7 was replaced by lines 25-29 |
- and then restart sics with the command |
+ |
+ ''Outdated information:'' |
+ |
+ If the motor is not running, it may be necessary to load the motor |
+ parameters with the el734_test program. Type |
Line 9 was replaced by line 31 |
- monit restart sicsserver |
+ el734_test -help |
Line 11 was replaced by lines 33-36 |
- on the unix prompt. |
+ for documentation. A parameter file can be found under /home/users/tascom/PSI/mot_4_01.par (no longer there!) |
+ |
+ Actually the serial cable for the omega motor is connected to channel 4 (port 3004) on the |
+ terminal server lnsts02.psi.ch. |