Line 1 was replaced by line 1 |
- !MA09 Specifications |
+ !!MA09 Specifications |
At line 2 added 5 lines. |
+ !Sample space dimensions |
+ see [Magnet List] |
+ |
+ !Other specifications |
+ |
Lines 12-14 were replaced by lines 17-19 |
- |at 1 m horizontal distance|235 Gauss |
- |at 1 m vertical distance|310 Gauss |
- |critical distance for loose ferromagnetic parts\\500 Gauss/metre|1.15 m vertical\\1.10 m horizontal |
+ |at 1 m horizontal distance|150 Gauss |
+ |at 1 m vertical distance|300 Gauss |
+ |critical distance for loose ferromagnetic parts\\500 Gauss/metre|0.84 m vertical\\0.67 m horizontal |
Line 21 was replaced by line 26 |
- |max. useable sample height|40 mm |
+ |max. beam height for horizontal access|40 mm |
Line 23 was replaced by line 28 |
- |sample height for +-6 deg access(visible by neutrons) |24 mm |
+ |max. beam height for +-6 deg access|24 mm |
Line 28 was replaced by line 33 |
- |total weight including cryogens|? kg |
+ |LHe hold time with stable temperature below 50 K\\(as of Feb 2005) | 2.9 days |
Line 30 was replaced by line 35 |
- |LHe boil off (static mode)|330 cc/hr |
+ |LHe boil off (static mode, original spec)|330 cc/hr |
At line 32 added 1 line. |
+ |total weight including cryogens|? kg |