Line 1 was replaced by line 1 |
- !MA09 Specifications |
+ !!MA09 Specifications |
At line 2 added 5 lines. |
+ !Sample space dimensions |
+ see [Magnet List] |
+ |
+ !Other specifications |
+ |
Line 6 was replaced by line 11 |
- |operating current for 9 T|< 120 A |
+ |operating current for 9 T|113.34 A |
Lines 12-14 were replaced by lines 17-19 |
- |at 1 m horizontal distance|235 Gauss |
- |at 1 m vertical distance|310 Gauss |
- |critical distance for loose ferromagnetic parts\\500 Gauss/metre|1.15 m vertical\\1.10 m horizontal |
+ |at 1 m horizontal distance|150 Gauss |
+ |at 1 m vertical distance|300 Gauss |
+ |critical distance for loose ferromagnetic parts\\500 Gauss/metre|0.84 m vertical\\0.67 m horizontal |
Lines 21-24 were replaced by lines 26-29 |
- |max. useable sample height|30 mm |
- |split access angle|+- 5 deg |
- |sample height for +-5 deg access(visible by neutrons) |? |
- |horizontal access angle|> 330 deg |
+ |max. beam height for horizontal access|40 mm |
+ |split access angle|+- 6 deg |
+ |max. beam height for +-6 deg access|24 mm |
+ |horizontal access angle|340 deg |
Removed line 26 |
- |entrance window size (radiussed corners 3 mm)|? x ? mm |
At line 28 added 5 lines. |
+ |LHe hold time with stable temperature below 50 K\\(as of Feb 2005) | 2.9 days |
+ |useable LHe capacity|45 l |
+ |LHe boil off (static mode, original spec)|330 cc/hr |
+ |LN2 capacity|43 l |
+ |typical LN2 boil off|630 cc/hr |
Removed lines 30-33 |
- |useable LHe capacity|? l |
- |LHe boil off (static mode)|? cc/hr |
- |LN2 capacity|? l |
- |typical LN2 boil off|? cc/hr |