| Lines 12-15 were replaced by lines 12-36 |
| - We have to modify sligtly the design of the radiation shields and establish |
| - the knowledge how to produce the radiation shields and how to assemble them. |
| - ''As we have actually no replacement parts, the furnace is operational only |
| - until the next damage''. |
| + The furnace is back from repair at ''AS Scientific Products Ltd'' and |
| + operational. |
| + The screens are now in a modified design, which allows an easier |
| + replacement of the heater element. But even with this new design, |
| + a repair will take a lot of time. |
| + |
| + !Power table |
| + |
| + When heating up the furnace the first time, ramp up slowly, using |
| + the appropriate max. power for the temperature range. The pid values |
| + from a first test (values in parentheses) were not confirm in a later |
| + experiment. |
| + |
| + ||temperature||max. power||prop||int||ramp |
| + |1500 ... 1800|2000|(50) 10|(50) 5|20 |
| + |1300 ... 1500|1000|(100)|(50)|20 |
| + |1100 ... 1300| 500|(200)|(50)|20 |
| + | 900 ... 1100| 316|(200)|(50)|10 |
| + | 700 ... 900| 158|(100)|(20)|5 |
| + | 600 ... 700| 158|(100)|(1)|5 |
| + | 500 ... 600| 158|(50)|(1)|5 |
| + |
| + [illofen1.jpg] |
| + [illofen2.jpg] |
| + [illofen4.jpg] |