Difference between version 42 and version 41:
At line 39 added 1 line.
+
Line 72 was replaced by lines 73-84
-
+ !Returning Liquid Helium Dewars
+ please return dewars to the dewar station as soon as possible after the transfer
+ *connect the dewar to the rubber hose of the dewar station
+ *open the manual valve to the recovery
+ *open the manual pressure release that is connected to the recovery on the dewar
+ *check the other manual pressure release. it MUST be closed
+ *retighten the locking nut of the Helium extraction port
+ *connect the level probe
+
+ IMPORTANT:
+ recheck that all valves are in the right state and the level is reading before you leave
+ [afterfill.png]
At line 73 added 1 line.
+ In case you are unsure about certain procedures, please contact sample environment staff