Difference between version 42 and version 3:
Line 3 was replaced by line 3
- !liquid Nitrogen
+ !General Remarks
Lines 5-8 were replaced by lines 5-19
- Most of our cryostats and cryo-magnets have an outer liquid nitrogen jacket to reduce the heat load on the helium bath.
- The storage dewars are prepared by the sample environment staff and can be picked up at the preparation area (between FOCUS and MARS).
- The dewars provided are self pressurizing systems with a mechanical level indicator at the top as shown below.
- [LN2_dewar_schematics|cl_clpb.gif]
+ Handling cryogenic liquids should only be done by trained people. The following instruction are by no means a replacement for a certified cryogenics course. They are meant as a reference for how to handle cryogenics at the Neutron guide hall. \\
+ Fear is a bad teacher
+ and you should always consult a more experienced instrument scientist or [the sample environment team|SampleEnvironment],
+ if you are in doubt how to handle things correctly. \\
+ Please always keep in mind that you are working with boiling liquids which will evaporate rather quickly when exposed to ambient temperatures.\\
+ Think about the consequences of your doing __before__ you do it (e.g. what will happen if you open a specific valve).
+
+ For information on how to set up an orange cryostat and how to change samples, please follow this link: [OrangeCryostat|http://lns00.psi.ch/sinqwiki/Wiki.jsp?page=OrangeCryostatCooldownProcedure].
+
+ !Liquid Nitrogen
+
+ Most of our cryostats and cryo-magnets have an outer liquid nitrogen jacket to reduce the heat load on the helium bath.\\
+ The storage dewars are prepared by the sample environment staff and can be picked up at the preparation area (between FOCUS and MARS).\\
+ The dewars provided are self pressurizing systems with a mechanical level indicator at the top as shown below.\\
+ [LN2_dewar_schematics|LN2.png]
Lines 12-13 were replaced by lines 23-24
- The filling hoses have a metallic protection sleeve. The valves and the hose get very cold during a nitrogen transfer.
- *Make sure that you wear protective leather gloves when handling cold parts.
+ The filling hoses have a metallic protection sleeve. The valves and the hose get very cold during a nitrogen transfer.
+ *Make sure that you wear protective leather gloves when handling cold parts, however the gloves should never be soaked with LN2 or LHe.
At line 14 added 1 line.
+
At line 17 added 10 lines.
+ !Liquid Helium
+
+ Liquid Helium dewars are provided by the sample environment staff at the preparation area.\\
+ At PSI we run a Helium recovery system.
+ *Please make sure that the dewars are always connected to the recovery line when not in use
+ *Remember to close the valve on the recovery line when taking a dewar away and to open it after return
+ *The Level meter in the preparation area shows the level in litres. Make sure that there is sufficient helium for filling. The dewars
+ should be returned with a level of at least 10-15 litres.
+ *When using the crane be aware that no equipment (or dewar) should be moved over the DMC detector
+
Line 21 was replaced by lines 43-47
- [He_gas_line|gasline.png]
+ __Helium Dewar Transport:__\\
+ Helium dewars are brought into the experimental area with the 1T crane.
+ *Only instructed people are allowed to use the crane. (A short training is offered by the LDM mechanics personal.)
+ *Only dewars with pre-installed transport straps are allowed for crane transportation.
+ *To extend the straps please use the proper extensions as shown in the picture below.
At line 22 added 1 line.
+ | [transport extensions|HeDewartransport.png] | [dewar with transport fixtures|transportstraps.png]
Line 24 was replaced by lines 51-86
- [CCU|CCU.png]
+
+ __Helium Transfer:__
+ *Before you open the extraction port you should release the pressure in the dewar which might have built up during transport by opening one of the release valves
+ *Insert the Helium siphon first into the transport dewar.
+ *close all manual pressure release valves
+ *slowly lower the siphon into the dewar and observe the pressure rise in the dewar (try to stay below 0.2bar)
+ *you should see gas streaming out on the other end of the siphon
+ *don't put your hands in the gas stream
+ *when it starts to turn white you can insert this end into the cryostat Helium filling port
+ *continue to lower the siphon into the transport dewar until you reach the bottom.
+ *switch on the compressed Helium gas line and connect the gas hose to one of the pressure release valves
+ and open it. The dewar is now pressurized to 0.15 bar over pressure.
+ *make sure not to empty the vessel completely, the people running the liquefier will love you. On the transfer line, normally there is a prolongation with a hole at 10 cm above the end. If not, you should lift the transfer line by 10 cm after inserting.
+ When the transfer is finished:
+
+ *stop the pressure line
+ *release the dewar pressure before you remove the siphon.
+ *make sure that the extraction and filling ports are properly sealed. It might be necessary to
+ use a hair dryer to remove the ice.
+
+ [He_gas_line|gasline.png]
+
+ !Returning Liquid Helium Dewars
+ please return dewars to the dewar station as soon as possible after the transfer
+ *connect the dewar to the rubber hose of the dewar station
+ *open the manual valve to the recovery
+ *open the manual pressure release that is connected to the recovery on the dewar
+ *check the other manual pressure release. it MUST be closed
+ *retighten the locking nut of the Helium extraction port
+ *connect the level probe
+
+ IMPORTANT:
+ recheck that all valves are in the right state and the level is reading before you leave
+ [afterfill.png]
+
+ In case you are unsure about certain procedures, please contact sample environment staff