NeedleValveController

Search SINQ Wiki:
SINQ LIN

SINQ Wiki
- Main Page
- Search SINQ Wiki
- Sample Environment
- Probenumg. Intern
- Troubleshooting SICS

This Page
- Page Info
- Printer Friendly

Referenced by
EarlierProgress

Wiki Info
- Unused pages
- Undefined pages
- RecentChanges
- Page Index
- System Info
- JSPWiki Docu
- SandBox
- OneMinuteWiki
- Create a New Page




JSPWiki v2.0.52


This is version 3. It is not the current version, and thus it cannot be edited.
[Back to current version]   [Restore this version]


Needle Valve Controller

The main component is an Eurotherm 2216e controller configured as a motorized valve controller. The box contains also a bipolar power supply used for the the pressure sensor and a network of resistors and diodes for the asymmetric power supply of the DC motor. The motor current is limited to 40/50 mA for closing/opening and may be reduced by means of a resistor in the motor box.

There is an upper end switch touched when the valve is sufficiently open, but there is no lower end switch. The motor is rated in a way that it does not damage when blocked.


Attachments:  NeedleValveMotor.jpg  NeedleValveControl.gif  NeedleValveController.jpg 
This particular version was published on 04-Aug-2006 09:07:35 UTC by MarkusZolliker.