| on cryostat | on rack (should not be disconnected) |
switch heater, warn light | (11) small 6 pin LEMO below warn light | switch heater on IPS (40) fixed connection to warn light power supply |
VTI sensors & heater | (12) 10 pin fisher on VTI | (4a/4b) sensor A/sensor B + heater (42) via relay (43) on LakeShore 340 |
LHe level | (13) small 4 pin (fisher?) on LHe3 neck | sensor 1 (4h) on ILM211 |
LN2 level | (21) large 4 pin (fisher?) on LN2 neck | sensor 2 (4n) on ILM211 |
sample sensors | (14) top of sample stick | (41c/41d) sensor C/sensor D on LakeShore 340 + heater OUT2 (44) |
VTI needle valve motor (not used) | leave unconnected | ITC503 15 pin DSUB on green platine |
He recovery | (16) He neck recovery port (without valve) | He recovery |
VTI pumping line | (17) VTI pumping port (do not open valve before tube is pumped out) | (31) VTI pump (top port) |
recovery of pumped He (VTI) | (32) VTI pump (side port) | He recovery |
lambda pump (for fields > 10 Tesla) | (18) lambda pump port on He neck | (31) Lambda pump (top port) |
recovery of pumped He (lambda) | (32) lambda pump (side port) | He recovery |