ILL3 is a cryofurnace which allows in principle temperatures from 2 K to 450 K without sample change. However, there are some drawbacks:

  • due to the large temperature range Rh-Fe sensors are used. The calibration at low temperatures is bad (about +- 0.3 K).
  • for high temperatures, one needs a special power supply for boosting the heating power.
  • for temperatures > 300 K the consumption of LHe (liquid helium) increases drastically
  • it is possible to work without LHe, under the condition, that the main vacuum is pumped constantly. The use of LN2 (liquid nitrogen) is mandatory.
  • the consuption of liquids may be reduced by removing the exchange gas. But this will affect the temperature of the sample, the difference to the sensor temperature might get untolerably high.

No Cadmium or Indium beyond Room Temperature!

LHe hold time with stable temperature below 50 K: 2.2 d

ill3a.jpg


Attachments:  ill3a.jpg 
This particular version was published on 12-Apr-2017 07:15:56 UTC by MarkusZolliker.