ILL3 is a cryofurnace which allows in principle temperatures from 2 K to
600 K without touching the sample. However, there are some drawbacks:
- due to the large temperature range Rh-Fe sensors are used. The calibration
at low temperatures is bad (about +- 0.3 K).
- for high temperatures, one needs a special power supply for boosting
the heating power.
- for temperatures > 300 K the consumption of LHe (liquid helium) increases
drastically
- it is possible to work without LHe, under the condition, that the main
vacuum is pumped constantly. The use of LN2 (liquid nitrogen) is mandatory.
- the consuption of liquids may be reduced by removing the exchange gas. But
this will affect the temperature of the sample, the difference to the sensor
temperature might get untolerably high.
LHe hold time with stable temperature below 50 K: | 2.2 d |