ILL3

Search SINQ Wiki:
SINQ LIN

SINQ Wiki
- Main Page
- Search SINQ Wiki
- Sample Environment
- Probenumg. Intern
- Troubleshooting SICS

This Page
- Page Info
- Printer Friendly

Referenced by
HeCryostats

Wiki Info
- Unused pages
- Undefined pages
- RecentChanges
- Page Index
- System Info
- JSPWiki Docu
- SandBox
- OneMinuteWiki
- Create a New Page




JSPWiki v2.0.52


This is version 1. It is not the current version, and thus it cannot be edited.
[Back to current version]   [Restore this version]


ILL3 is a cryofurnace which allows in principle temperatures from 2 K to 600 K without touching the sample. However, there are some drawbacks:

  • for high temperatures, one needs a special power supply for boosting the heating power.
  • for temperatures > 300 K the consumption of LHe (liquid helium) increases drastically
  • it is possible to work without LHe, under the condition, that the main vacuum is pumped constantly. The use of LN2 (liquid nitrogen) is mandatory.
  • the consuption of liquids may be reduced by removing the exchange gas. But this will affect the temperature of the sample, the difference to the sensor temperature might get untolerably high.


Attachments:  ill3a.jpg 
This particular version was published on 11-Feb-2005 14:06:50 UTC by MarkusZolliker.