This is version 3.
It is not the current version, and thus it cannot be edited.
[Back to current version]
[Restore this version]
Oxford Intruments 3He Sorption Cryostat
HEOX is a 3He cryostat with a sorption 3He pump. All gas handling is
inside the cryostat, there are therefore no tubes to a pump outside.
After putting into operation, some questions are still open:
- The sample space has a diameter of about 40 mm. Around this tube,
there is 4.5 mm liquid helium. This may produce background,
especially on inelastic measurements.
- In principle, it should be possible to combine the 3He insert with
the MA09 cryomagnet. It is not yet clear, if it really works, and
there will also be the problem with the LHe background.
- For powders, it is not clear if the thermal contact inside the
sample container is good enough to ensure the expected sample
temperature.
Some test results
Hold times:
0.33 K | 22h |
0.5 K | 25h |
0.7 K | 32h |
Attachments:
he3ontasp.gif
helium3a.jpg
This particular version was published on 12-Apr-2005 06:16:53 UTC by MarkusZolliker.
|