IPS

Oxford Instruments superconducting magnet power supply IPS

Create an IPS object:

  makeobject obj mf ips <terminal server> <port>
The name might be choosen differently, mf is used by convention, and is also used by the scripts called with the samenv command.

Subcommands and Parameters

Command and subcommand Description
mf rmagnetic field
mf persmode w0: do not work in persistent mode, 1: go to persistent mode after a drive
mf perswitch rstate of persistent mode switch
mf maxlimit rmaximal limit for field
mf limit wsoft limit for field (when set the first time after initialization, maxlimit is also set)
mf ramp wramp rate (the real ramp rate may be limited by the sweep limits defined in the IPS)
mf current rthe current in the leads (converted to Tesla)
mf lastfield rvalue the the field before goings to persistent mode
mf confirm cconfirm the persistent field in case the stored value in the driver does not match the value stored in the IPS
mf targetValuertarget value

r: read-only w: writeable parameter

Plus standard commands and parameters defined by EaseBasePar, EaseSendPar, EaseMsgPar