Diff: MagnetList

Search SINQ Wiki:
SINQ LIN

SINQ Wiki
- Main Page
- Search SINQ Wiki
- Sample Environment
- Probenumg. Intern
- Troubleshooting SICS

This Page
- Page Info
- Printer Friendly

Referenced by
MA6
MA11Specifications
MA09Specifications
MA15Specifications
MA7Specifications
SampleEnvironment

Wiki Info
- Unused pages
- Undefined pages
- RecentChanges
- Page Index
- System Info
- JSPWiki Docu
- SandBox
- OneMinuteWiki
- Create a New Page




JSPWiki v2.0.52


Difference between version 104 and version 71:
At line 2 added 1 line.
+ The sample mounting thread on all sticks is M6 (female).\\
Removed line 5
- Our standard thread on all systems is M6 female.
Lines 7-14 were replaced by lines 7-15
- ||Name||Description||Field||Temp.||d\\(without\\[dilution|DilSampleMounting])||a\\[[mm]||b\\[[mm]||l\\[[mm]||g\\[[mm]||h\\[[mm]||200 gauss\\line[[m]
- |[MA15]|Oxford Instr. Cryomagnet |vert. 14.9 T |1.6...300 K\\>80 mK[1] |20 |14...90 |139 |1294 (__1454__)|49.5|1078(+2)|1.89
- |[MA11]|Oxford Instr. SANS Cryomagnet|hor. 11 T |1.8...300 K\\>80 mK[1] |22 |75...100[3]|54|1302 (__1462__)|62 |1080(+2)|
- |[MA09]|Oxford Instr. Cryomagnet |vert. 9 T |1.8...300 K\\>80 mK[1] |48 |93...137[3]|58|__1247__ |55 | |1.14
- |[MA7] |Oxford Instr. Cryomagnet |hor. 6.8 T |1.8...300 K\\>80 mK[1] |48 |35...90 |69 |1357 (__1517__)|49.5| |1.19
- |[MA6] |Oxford Instr. Cryomagnet |vert. 6 T |1.8...300 K\\>80 mK[1] |30 |40...90 |80 |__1309__ (__1469__{{*}})|49.5|1078(+2)|0.70
- |[MA02]|Oxford Instr. Cryomagnet |hor. 1.8 T |1.8...150 K[4]\\>100 mK[1] |28 | 0...56 |93 |1245 (1405) |55 |968(+2) |
- |[MA4] |Thor Cryomagnet |vert. 4 T |4.2...100 K |24.5|37...71|29 |__1286__ |24.5| |
+ ||Name||Description||Field||Temp.||d\\(without\\[dilution|DilSampleMounting]) ||a\\[[mm]||b\\[[mm]|| L\\[[mm]||g\\[[mm]||h\\[[mm]
+ |[MA15]|Oxford Instr. Cryomagnet |vert. 14.9 T |1.6...300 K\\>80 mK[1] |20 |14...90 | 139| __1454__ |49.5|1238(+2)
+ |[MA11]|Oxford Instr. SANS Cryomagnet|hor. 11 T |1.8...300 K\\>80 mK[1] |22 |75...100[3]|54| __1462__ |49.5 (62)|1240(+2)
+ |[MA7] |Oxford Instr. Cryomagnet |hor. 6.8 T |1.8...300 K\\>80 mK[1] |48 |35...90 | 69| __1517__ |49.5|
+ |[MA6] |Oxford Instr. Cryomagnet |vert. 6 T |1.8...300 K\\>80 mK[1] |30 |40...90 | 80| __1469__ |49.5|1078(+2)
+ |[MA02]|Oxford Instr. Cryomagnet |hor. 1.8 T |1.8...150 K[4]\\>100 mK[1] |28 | 0...56 | 93| __1405__ |55 |968(+2)
+ |[MA10]|Oxford Instr. Cryomagnet |vert. 10 T |1.8...300 K\\>80 mK[1] |48 | | 115| __1275__ |49.5|
+ |[MB11]|Oxford Instr. Cryomagnet |vert. 11T |1.8..300K\\>50mK[1] |100|50...150| 150| __1551__[2] |100 |
+ |''[MA09]''|''out of service'' |vert. 9 T |1.6...320 K\\>80 mK[1] |48 |93...137[3]|58| __1247__ |55 |
At line 16 added 1 line.
+ [dimensions|HeCryostats/ORIdimensions.png]\\
Line 20 was replaced by line 22
- l: length from top flange to the center of the sample (when used with a rotation unit, add 160)\\
+ L: length from top flange to the center of the sample, including rotation stage (when used without, reduce by 160). Please be aware that the values given are to measured at RT, a 2-3 mm shrinking at low T is already considered.\\
Removed line 23
- {{*}}: MA6 with rotation, measured on [TriCS|http://www.psi.ch/sinq/trics/trics] (geometrical: 1464)\\
At line 26 added 1 line.
+ [2|#2] the rotation stage can not be removed\\
Lines 32-35 were replaced by line 34
- !!!Stray fields
- The stray fields can be calculated from the 200 gauss line distance.
- For vertical magnets, this is the vertical distance. At the same distance in horizontal direction, the stray field is 100 gauss (0.01 Tesla).
- The stray field folows a r^-3 rule, what means that at half the distance you have a 8 times higher field.
+ [Strayfields]

Back to MagnetList, or to the Page History.