Difference between version 104 and version 40:
Line 3 was replaced by lines 3-15
- Please read also the [rules concerning superconducting magnet experiments|MagnetDilutionPolicy]
+ The sample mounting thread on all sticks is M6 (female).\\
+ Please read also the [rules concerning superconducting magnet experiments|MagnetDilutionPolicy]\\
+ and the [Introduction on cryogenic handling |http://lns00.psi.ch/sinqwiki/Wiki.jsp?page=CryogenicHandling]\\
+
+ ||Name||Description||Field||Temp.||d\\(without\\[dilution|DilSampleMounting]) ||a\\[[mm]||b\\[[mm]|| L\\[[mm]||g\\[[mm]||h\\[[mm]
+ |[MA15]|Oxford Instr. Cryomagnet |vert. 14.9 T |1.6...300 K\\>80 mK[1] |20 |14...90 | 139| __1454__ |49.5|1238(+2)
+ |[MA11]|Oxford Instr. SANS Cryomagnet|hor. 11 T |1.8...300 K\\>80 mK[1] |22 |75...100[3]|54| __1462__ |49.5 (62)|1240(+2)
+ |[MA7] |Oxford Instr. Cryomagnet |hor. 6.8 T |1.8...300 K\\>80 mK[1] |48 |35...90 | 69| __1517__ |49.5|
+ |[MA6] |Oxford Instr. Cryomagnet |vert. 6 T |1.8...300 K\\>80 mK[1] |30 |40...90 | 80| __1469__ |49.5|1078(+2)
+ |[MA02]|Oxford Instr. Cryomagnet |hor. 1.8 T |1.8...150 K[4]\\>100 mK[1] |28 | 0...56 | 93| __1405__ |55 |968(+2)
+ |[MA10]|Oxford Instr. Cryomagnet |vert. 10 T |1.8...300 K\\>80 mK[1] |48 | | 115| __1275__ |49.5|
+ |[MB11]|Oxford Instr. Cryomagnet |vert. 11T |1.8..300K\\>50mK[1] |100|50...150| 150| __1551__[2] |100 |
+ |''[MA09]''|''out of service'' |vert. 9 T |1.6...320 K\\>80 mK[1] |48 |93...137[3]|58| __1247__ |55 |
Removed lines 5-15
- Our standard thread on all systems is M6 female.
-
- ||Name||Description ||Field ||Temp. ||d ||a ||b ||l
- |[MA15]|Oxford Instr. Cryomagnet |vert. 14.9 T |1.6...300 K\\>80 mK[2] |20 |14...90 |139 |1294+160=1454
- |[MA11]|Oxford Instr. SANS Cryomagnet|hor. 11 T |1.8...300 K\\>80 mK[2] |22 |75...100[3]|54|1302+160=1463
- |[MA09]|Oxford Instr. Cryomagnet |vert. 9 T |1.8...300 K\\>80 mK[1] |48 |93...137[3]|58|1247
- |[MA7] |Oxford Instr. Cryomagnet |hor. 6.8 T |1.8...300 K\\>80 mK[1] |48 | | |
- |[MA6] |Oxford Instr. Cryomagnet |vert. 6 T |1.8...300 K\\>80 mK[1] |30 |40...90 |80 |1304
- |[MA02]|Oxford Instr. Cryomagnet |hor. 1.8 T |1.8...150 K[4]\\>100 mK[1] |28 | 0...56 |93 |1245
- |[MA4] |Thor Cryomagnet |vert. 4 T |4.2...100 K |24.5|37...71|29 |1286
-
At line 16 added 1 line.
+ [dimensions|HeCryostats/ORIdimensions.png]\\
Line 20 was replaced by lines 22-25
- l: length from top flange to the center of the sample (when used with a rotation unit, add 160)\\
+ L: length from top flange to the center of the sample, including rotation stage (when used without, reduce by 160). Please be aware that the values given are to measured at RT, a 2-3 mm shrinking at low T is already considered.\\
+ g: diameter of upper part of sample stick space\\
+ h: length of upper (larger) part of sample stick space (+2mm when warm)\\
+ The allowed sample diameter in combination with a dilution can be found [here|DilSampleMounting]\\
Lines 22-23 were replaced by lines 27-28
- [1|#1] in combination with dilution sticks [DIL1] & [DIL2]\\
- [2|#2] in combination with dilution stick [DIL2]\\
+ [1|#1] in combination with dilution sticks\\
+ [2|#2] the rotation stage can not be removed\\
At line 28 added 2 lines.