- Please read also the [rules concerning superconducting magnet experiments|MagnetDilutionPolicy]
+ The sample mounting thread on all sticks is M6 (female).\\
+ Please read also the [rules concerning superconducting magnet experiments|MagnetDilutionPolicy]\\
+ and the [Introduction on cryogenic handling |http://lns00.psi.ch/sinqwiki/Wiki.jsp?page=CryogenicHandling]\\
+
+ ||Name||Description||Field||Temp.||d\\(without\\[dilution|DilSampleMounting]) ||a\\[[mm]||b\\[[mm]|| L\\[[mm]||g\\[[mm]||h\\[[mm]
+ |[MA15]|Oxford Instr. Cryomagnet |vert. 14.9 T |1.6...300 K\\>80 mK[1] |20 |14...90 | 139| __1454__ |49.5|1238(+2)
+ |[MA11]|Oxford Instr. SANS Cryomagnet|hor. 11 T |1.8...300 K\\>80 mK[1] |22 |75...100[3]|54| __1462__ |49.5 (62)|1240(+2)
+ |[MA7] |Oxford Instr. Cryomagnet |hor. 6.8 T |1.8...300 K\\>80 mK[1] |48 |35...90 | 69| __1517__ |49.5|
+ |[MA6] |Oxford Instr. Cryomagnet |vert. 6 T |1.8...300 K\\>80 mK[1] |30 |40...90 | 80| __1469__ |49.5|1078(+2)
+ |[MA02]|Oxford Instr. Cryomagnet |hor. 1.8 T |1.8...150 K[4]\\>100 mK[1] |28 | 0...56 | 93| __1405__ |55 |968(+2)
+ |[MA10]|Oxford Instr. Cryomagnet |vert. 10 T |1.8...300 K\\>80 mK[1] |48 | | 115| __1275__ |49.5|
+ |[MB11]|Oxford Instr. Cryomagnet |vert. 11T |1.8..300K\\>50mK[1] |100|50...150| 150| __1551__[2] |100 |
+ |''[MA09]''|''out of service'' |vert. 9 T |1.6...320 K\\>80 mK[1] |48 |93...137[3]|58| __1247__ |55 |
Removed lines 5-11
- ||Name||Description ||Field ||Temp. ||d ||a ||b||Thread
- |[MA15]|Oxford Instr. Cryomagnet |vert. 14.9 T |1.6...300 K\\>80 mK[2] |20 |14...90 |139|M6
- |[MA11]|Oxford Instr. SANS Cryomagnet|hor. 11 T |1.8...300 K\\>80 mK[2] |22 |75...100[3]|44|M8 (or M6 with adapter)
- |[MA09]|Oxford Instr. Cryomagnet |vert. 9 T |1.8...300 K\\>80 mK[1] |48 |93...137[3]|58|M20x1 (or M6[5])
- |[MA02]|Oxford Instr. Cryomagnet |hor. 1.8 T |1.8...150 K[4]\\>100 mK[1] |28 | 0...56 |93|M20x1 (or M6[5])
- |[MA4] |Thor Cryomagnet |vert. 4 T |4.2...100 K |24.5|37...71|29|M6
-
At line 12 added 1 line.
+ [dimensions|HeCryostats/ORIdimensions.png]\\
Line 16 was replaced by lines 22-25
- M6,M8,M20x1: a hole on the stick with the specified thread
+ L: length from top flange to the center of the sample, including rotation stage (when used without, reduce by 160). Please be aware that the values given are to measured at RT, a 2-3 mm shrinking at low T is already considered.\\
+ g: diameter of upper part of sample stick space\\
+ h: length of upper (larger) part of sample stick space (+2mm when warm)\\
+ The allowed sample diameter in combination with a dilution can be found [here|DilSampleMounting]\\
Lines 18-19 were replaced by lines 27-28
- [1|#1] in combination with dilution sticks [DIL1] & [DIL2]\\
- [2|#2] in combination with dilution stick [DIL2]\\
+ [1|#1] in combination with dilution sticks\\
+ [2|#2] the rotation stage can not be removed\\
Line 24 was replaced by lines 33-34
- [5|#5] adapters to M6 available and recommended
+