Line 3 was replaced by lines 3-15 |
- Please read also the [rules concerning superconducting magnet experiments|MagnetDilutionPolicy] |
+ The sample mounting thread on all sticks is M6 (female).\\ |
+ Please read also the [rules concerning superconducting magnet experiments|MagnetDilutionPolicy]\\ |
+ and the [Introduction on cryogenic handling |http://lns00.psi.ch/sinqwiki/Wiki.jsp?page=CryogenicHandling]\\ |
+ |
+ ||Name||Description||Field||Temp.||d\\(without\\[dilution|DilSampleMounting]) ||a\\[[mm]||b\\[[mm]|| L\\[[mm]||g\\[[mm]||h\\[[mm] |
+ |[MA15]|Oxford Instr. Cryomagnet |vert. 14.9 T |1.6...300 K\\>80 mK[1] |20 |14...90 | 139| __1454__ |49.5|1238(+2) |
+ |[MA11]|Oxford Instr. SANS Cryomagnet|hor. 11 T |1.8...300 K\\>80 mK[1] |22 |75...100[3]|54| __1462__ |49.5 (62)|1240(+2) |
+ |[MA7] |Oxford Instr. Cryomagnet |hor. 6.8 T |1.8...300 K\\>80 mK[1] |48 |35...90 | 69| __1517__ |49.5| |
+ |[MA6] |Oxford Instr. Cryomagnet |vert. 6 T |1.8...300 K\\>80 mK[1] |30 |40...90 | 80| __1469__ |49.5|1078(+2) |
+ |[MA02]|Oxford Instr. Cryomagnet |hor. 1.8 T |1.8...150 K[4]\\>100 mK[1] |28 | 0...56 | 93| __1405__ |55 |968(+2) |
+ |[MA10]|Oxford Instr. Cryomagnet |vert. 10 T |1.8...300 K\\>80 mK[1] |48 | | 115| __1275__ |49.5| |
+ |[MB11]|Oxford Instr. Cryomagnet |vert. 11T |1.8..300K\\>50mK[1] |100|50...150| 150| __1551__[2] |100 | |
+ |''[MA09]''|''out of service'' |vert. 9 T |1.6...320 K\\>80 mK[1] |48 |93...137[3]|58| __1247__ |55 | |
Removed lines 5-11 |
- ||Name||Description ||Field ||Temp. ||d ||a ||b||Thread |
- |[MA15]|Oxford Instr. Cryomagnet |vert. 14.9 T |1.6...300 K |20 |14...90 |139|M6 |
- |[MA11]|Oxford Instr. SANS Cryomagnet|hor. 11 T |1.8...300 K |22 | | |M8 |
- |[MA09]|Oxford Instr. Cryomagnet |vert. 9 T |1.8...300 K\\>100 mK[1]|48 |93...137[2]|58|M20x1 |
- |[MA02]|Oxford Instr. Cryomagnet |hor. 1.8 T |1.8...150 K [3] |28 | 0...56 |93|M20x1 |
- |[MA4] |Thor Cryomagnet |vert. 4 T |4.2...100 K |24.5|37...71|29|M6 |
- |
At line 12 added 1 line. |
+ [dimensions|HeCryostats/ORIdimensions.png]\\ |
Lines 16-19 were replaced by lines 22-29 |
- M6,M8,M20x1: a hole on the stick with the specified thread |
- |
- [1|#1] in combination with dilution stick\\ |
- [2|#2] as we have adapter pieces, it is recommended to make the sample |
+ L: length from top flange to the center of the sample, including rotation stage (when used without, reduce by 160). Please be aware that the values given are to measured at RT, a 2-3 mm shrinking at low T is already considered.\\ |
+ g: diameter of upper part of sample stick space\\ |
+ h: length of upper (larger) part of sample stick space (+2mm when warm)\\ |
+ The allowed sample diameter in combination with a dilution can be found [here|DilSampleMounting]\\ |
+ |
+ [1|#1] in combination with dilution sticks\\ |
+ [2|#2] the rotation stage can not be removed\\ |
+ [3|#3] as we have adapter pieces, it is recommended to make the sample |
Lines 21-22 were replaced by lines 31-34 |
- [3|#3] for a short time 300 K is possible, but not recommended due |
- to very high liquid He consumption |
+ [4|#4] for a short time 300 K is possible, but not recommended due |
+ to very high liquid He consumption\\ |
+ |
+ [Strayfields] |