Diff: MA15TASPRestrictions

Search SINQ Wiki:
SINQ LIN

SINQ Wiki
- Main Page
- Search SINQ Wiki
- Sample Environment
- Probenumg. Intern
- Troubleshooting SICS

This Page
- Page Info
- Printer Friendly

Referenced by
MA15

Wiki Info
- Unused pages
- Undefined pages
- RecentChanges
- Page Index
- System Info
- JSPWiki Docu
- SandBox
- OneMinuteWiki
- Create a New Page




JSPWiki v2.0.52


Difference between version 15 and version 5:
Line 3 was replaced by line 3
- The operation on TASP is allowed __up to a field of 12 Tesla__ for A2 angles between -45 and -100.
+ The operation on TASP is allowed __up to a field of 14.9 Tesla__ for A2 angles between -45 and -100.
Line 12 was replaced by lines 12-35
- At 8.6 Tesla, A2=-48: vertical force 440 N, horizontal force 210 N.
+ Summary: Distance as above, at 8.6 Tesla, a2=-48, a4=105: vertical force 440 N, horizontal force 210 N.\\
+ [Report by Martin Mansson|MA15TASPRestrictions/TASP_M15_forcetest.pdf]
+
+ !Force test 27. July 2015
+
+ Conditions:
+ * Ei =11meV == a2 = -47.97deg
+ * a4 = 105deg
+ * Be-filter installed
+ * dist. Magnet-filter: 260mm
+ * dist. mono-table: 630mm
+
+ Both vertical and horizontal forces stay below the specified maximum allowance.
+ In this configuration the magnet can be used up to 14.9T.
+
+ [vertiforce.png]
+
+ [horiforce.png]
+
+ [foto1.jpg]
+
+ [foto2.jpg]
+
+ [foto3.jpg]

Back to MA15TASPRestrictions, or to the Page History.