Lines 1-3 were replaced by line 1 |
- [MA11Installation/ma11_fig1.jpg] |
- [MA11Installation/ma11_fig2.jpg] |
- [MA11Installation/ma11_fig3.jpg] |
+ !Electrical Connections (these informations are partially obsolete) |
Removed lines 5-6 |
- !Electrical Connections |
- |
Line 8 was replaced by line 4 |
- |switch heater, warn light|(11) small 6 pin LEMO below warn light |switch heater on IPS (40)\\fixed connection to warn light power supply |
+ |switch heater, warn light|(11) small 6 pin LEMO below warn light |switch heater on IPS (45)\\fixed connection to warn light power supply |
Lines 10-12 were replaced by lines 6-9 |
- |LHe level |(13) small 4 pin (fisher?) on LHe3 neck|sensor 1 (4h) on ILM211 |
- |LN2 level |(21) large 4 pin (fisher?) on LN2 neck |sensor 2 (4n) on ILM211 |
- |sample sensors |(14) top of sample stick |(41c/41d) ''sensor C/sensor D'' on LakeShore 340 + heater ''OUT2'' (44) |
+ |Magnet power leads |(22) magnet power connections on LHe neck|(48) magnet power leads on IPS |
+ |LHe level |(13) small 4 pin (fisher?) on LHe neck |(4h) sensor 1 on ILM211 |
+ |LN2 level |(21) large 4 pin (fisher?) on LN2 neck |(4n) sensor 2 on ILM211 |
+ |sample sensors |(14) top of sample stick |(4c/4d) ''sensor C/sensor D'' on LakeShore 340 + heater ''OUT2'' (44) |
Lines 25-33 were replaced by lines 22-31 |
- | |SANS-I |SANS-II |AMOR |
- |LakeShore 340\\''SERIAL-I/O'' (41) |port 11\\RJ11 in grey adapter + gender changer|port 6 |port 10 |
- |IPS magnet power supply\\''RS232'' (46) with grey adapter and RJ45|port 12\\RJ45 in grey adapter + gender changer|port 7 | |
- |ILM211\\''RS232'' (40) with grey adapter and RJ45| | |
- !Tecs Configuration |
- |
- * the TecsServer must be running (if not: in the instruments account, type ''monit start tecs'') |
- * within a SICS client type ''tecs on'' and then ''tt device ma11''\\''ma11'' may be replaced by ''ma11r'' (for |
- the rotatable sample stick) or ''ma11a'' (for asgers special sample stick) |
+ | |SANS-I\\extra\\terminal server\\psts228|SANS-II |AMOR |
+ |LakeShore 340\\''SERIAL-I/O'' (41) |port 6|port 6 |port 10 |
+ |IPS magnet power supply\\''RS232'' (46) with grey adapter and RJ45|port 8|port 7 | |
+ |ILM211\\''RS232'' (40) with grey adapter and RJ45|port 9| |
+ |
+ [MA11Installation/ma11_fig1.jpg] |
+ [MA11Installation/ma11_fig2.jpg] |
+ [MA11Installation/ma11_fig3.jpg] |
+ [MA11Installation/ma11_fig4.jpg] |
+ |
Line 35 was replaced by line 33 |
- !Magnet power supply configuration |
+ !Software Configuration |
Line 37 was replaced by lines 35-40 |
- * within a SICS client type ''ips on'' and then ''ips help'' |
+ * the TecsServer and the SeaServer must be running |
+ (if not: in the instruments account, type |
+ ''monit start tecs'' / ''monit start sea'') |
+ * within SEA client select device |
+ ''ma11'' (for the original sample stick) or |
+ ''ma11r'' (for the rotatable sample stick) |