Line 1 was replaced by line 1 |
- !Electical Connections |
+ !Electrical Connections (these informations are partially obsolete) |
Lines 3-17 were replaced by lines 3-40 |
- || ||on cryostat ||on rack (should not be disconnected) |
- |switch heater, warn light|small 6 pin LEMO below warn light |switch heater on IPS\\fixed connection to warn light power supply |
- |VTI sensors & heater |10 pin fisher on VTI |sensors A & B + heater (via relay) on LakeShore 340 |
- |LN2 level |large 4 pin (fisher?) on LN2 neck |sensor 2 on ILM |
- |LHe level |small 4 pin (fisher?) on LHe3 neck|sensor 1 on ILM |
- |sample sensors |top of sample stick |sensors C & D on LakeShore 340 |
- |VTI needle valve motor (not used) |leave unconnected |ITC503 15 pin DSUB |
- |
- !Cryogenic connections |
- |
- |He recovery |He neck recovery port KF 25 (without valve) |He recovery |
- |VTI pumping line |VTI pumping port \\ (do not open valve before tube is pumped out) |VTI pump (top port) |
- |recovery of pumped He (VTI) |VTI pump (side port) |He recovery |
- |lambda pump (for fields > 10 Tesla)|lambda pump port on He neck |Lambda pump (top port) |
- |recovery of pumped He (lambda) |lambda pump (side port) |He recovery |
+ || ||on cryostat ||on rack (should not be disconnected) |
+ |switch heater, warn light|(11) small 6 pin LEMO below warn light |switch heater on IPS (45)\\fixed connection to warn light power supply |
+ |VTI sensors & heater |(12) 10 pin fisher on VTI |(4a/4b) ''sensor A/sensor B'' + heater (42) via relay (43) on LakeShore 340 |
+ |Magnet power leads |(22) magnet power connections on LHe neck|(48) magnet power leads on IPS |
+ |LHe level |(13) small 4 pin (fisher?) on LHe neck |(4h) sensor 1 on ILM211 |
+ |LN2 level |(21) large 4 pin (fisher?) on LN2 neck |(4n) sensor 2 on ILM211 |
+ |sample sensors |(14) top of sample stick |(4c/4d) ''sensor C/sensor D'' on LakeShore 340 + heater ''OUT2'' (44) |
+ |VTI needle valve motor (not used) |leave unconnected |ITC503 15 pin DSUB on green platine |
+ |
+ !Cryogenic Connections |
+ |
+ |He recovery |(16) He neck recovery port (without valve) |He recovery |
+ |VTI pumping line |(17) VTI pumping port \\ (do not open valve before tube is pumped out) |(31) VTI pump (top port) |
+ |recovery of pumped He (VTI) |(32) VTI pump (side port) |He recovery |
+ |lambda pump (for fields > 10 Tesla)|(18) lambda pump port on He neck |(31) Lambda pump (top port) |
+ |recovery of pumped He (lambda) |(32) lambda pump (side port) |He recovery |
+ |
+ !Computer Connections |
+ |
+ | |SANS-I\\extra\\terminal server\\psts228|SANS-II |AMOR |
+ |LakeShore 340\\''SERIAL-I/O'' (41) |port 6|port 6 |port 10 |
+ |IPS magnet power supply\\''RS232'' (46) with grey adapter and RJ45|port 8|port 7 | |
+ |ILM211\\''RS232'' (40) with grey adapter and RJ45|port 9| |
+ |
+ [MA11Installation/ma11_fig1.jpg] |
+ [MA11Installation/ma11_fig2.jpg] |
+ [MA11Installation/ma11_fig3.jpg] |
+ [MA11Installation/ma11_fig4.jpg] |
+ |
+ |
+ !Software Configuration |
+ |
+ * the TecsServer and the SeaServer must be running |
+ (if not: in the instruments account, type |
+ ''monit start tecs'' / ''monit start sea'') |
+ * within SEA client select device |
+ ''ma11'' (for the original sample stick) or |
+ ''ma11r'' (for the rotatable sample stick) |