Line 1 was replaced by line 1 |
- !!Switch heater test with MA09 |
+ !MA09 Switch Heater Test Results |
Line 3 was replaced by lines 3-7 |
- !Results: |
+ The switch heater output is specified 0-100 mA at 100 Ohm, which gives |
+ a maximum heater power of 1 W. The MA09 switch heater resistivity is |
+ 205 Ohm. As the current source has a limited voltage (17 V), the full current |
+ over the heater is no longer 100 mA, but only 83 mA. However, together with |
+ the increased resistivity, we get an maximum power of 1.41 W. |
Line 5 was replaced by line 9 |
- * Without lamda stage (coil at 4.2 K), a heater power of 0.81 W (13 V, 63 mA) |
+ * Without lambda stage (coil at 4.2 K), a heater power of 0.81 W (13 V x 63 mA) |
Line 7 was replaced by line 11 |
- * With lamda stage (coil at 2.3 K), a heater power of 1.06 W (14.7 V, 72 mA) |
+ * With lambda stage (coil at 2.3 K), a heater power of 1.06 W (14.7 V x 72 mA) |
Lines 9-11 were replaced by lines 13-15 |
- * The maximum switch heater power (205 Ohm switch heater) is 1.41 W (17 V x 83 mA). |
- * With the coil at 2.3 K it takes about 5 seconds from starting the heater until |
- the switch goes off (therefore the usual 30 sec. delay is very conservative). |
+ * With the coil at 2.3 K it takes about 5 seconds from starting the heater at maximum |
+ power (1.41 W) until the switch goes off (therefore the usual 30 sec. delay is very |
+ conservative). |