Line 5 was replaced by lines 5-8 |
- Please refer to the cryostat and magnet general assembly drawing in conjunction with the diagrams above. |
+ !''Symmetric Energisation (8/9 Tesla)'' |
+ Coaxial current lead terminal on tall cryostat services neck (this neck also features the VTI needle valve): |
+ * Inner terminal post I+ (start of entire magnet, internal red wire) |
+ * Outer terminal post I- (end of entire magnet, internal black wire) |
Line 8 was replaced by lines 11-14 |
- ''2.1.1 Magnet Service Necks'' |
+ !''Asymmetric Energisation (4 Tesla)'' |
+ Co-axial current lead terminal on short cryostat services neck (this neck also features the syphon entry port): |
+ * Inner terminal post I+ (start of selected coils, internal blue wire) |
+ * Outer terminal post I+ (end of selected coils, internal brown wire) |
Lines 10-29 were replaced by lines 16-18 |
- SYMMETRIC ENERGISATION (8/9 TESLA) |
- |
- Coaxial current lead terminal on tall cryostat services neck |
- (this neck also features the VTI needle valve): |
- lnner terminal post l+ (start of entire magnet) |
- (red wire from magnet) |
- Outer terminal post 1- (end of entire magnet) |
- (black wire from magnet) |
- |
- |
- ASYMMETRIC ENERGISATION (4 TESLA) Co-axial current lead terminal on short cryostat services neck (this neck also features the syphon entry port):lnner terminal post l+ (start of selected coils) |
- |
- (blue wire from magnet) |
- |
- Outer terminal post 1+ (end of selected coils) |
- (brown wire from magnet) |
- |
- Ten pin connector on services neck with syphon entry :Connections to persistent mode switch heaters. |
- |
- Ten pin connector on services neck with lambda point refrigerator needle valve and pumping line:Connections to carbon resistors for monitoring lambda point refrigerator. |
+ !''Connections to persistent mode switch heaters'' |
+ Ten pin connector on services neck with syphon entry: |
+ [MA09WiringManual/SwitchHeaterPlug.gif] |
Line 31 was replaced by lines 20-22 |
- Ten pin connector on services neck with VTI needle valve :Connections to helium level probe. |
+ !''Connections to carbon resistors for monitoring lambda point refrigerator'' |
+ Ten pin connector on services neck with lambda point refrigerator needle valve and pumping line: |
+ [MA09WiringManual/AllenBradleyPlug.gif] |
At line 32 added 2 lines. |
+ !''Connections to helium level probe.'' |
+ Ten pin connector on services neck with VTI needle valve |
Line 34 was replaced by lines 27-29 |
- VTI WIRING. |
+ !''VTI heat exchanger thermometry and non-inductive heater'' |
+ Ten pin connectors on VTI top plate: |
+ [MA09WiringManual/VTIPlug.gif] |
Line 36 was replaced by lines 31-33 |
- Ten pin connectors on VTI top plate :VTI heat exchanger thermometry and non-inductive heater. |
+ !''Heater and carbon resistor for clearing VTI needle valve blockages'' |
+ Ten pin connector on cryostat top plate: |
+ [MA09WiringManual/NeedleValvePlug.gif] |
Line 38 was replaced by lines 35-36 |
- Ten pin connector on cryostat top plate :Heater and carbon resistor for clearing VTI needle valve blockages. |
+ !''Auto-needle valve control for the VTI'' |
+ Not mounted (Five pin connector on services neck with motor drive) |
Line 40 was replaced by lines 38-39 |
- Five pin connector on services neck with motor drive :Auto-needle valve control for the VTI. |
+ !''Original Risų Thermometry Cabling'' |
+ [MA09WiringManual/OriginalRisoeCables.gif] |
Lines 45-51 were replaced by line 44 |
- Please see circuit diagram. |
- |
- [MA09WiringManual/AllenBradleyPlug.gif] |
- [MA09WiringManual/SwitchHeaterPlug.gif] |
- [MA09WiringManual/VTIPlug.gif] |
- [MA09WiringManual/NeedleValvePlug.gif] |
- [MA09WiringManual/OriginalRisoeCables.gif] |
+ [MA09WiringManual/InternalWiring.gif] |