| Line 45 was replaced by line 45 |
| - !!Sea/SICS COmmands |
| + !!Sea/SICS Commands |
| At line 48 added 3 lines. |
| + * disable the sample heater |
| + {{ts set 0}} |
| + |
| Lines 55-56 were replaced by line 58 |
| - * wait 15 min then stop heating the sorb |
| - |
| + * wait 15 min, then disable sorb heating |
| Line 59 was replaced by line 61 |
| - !regulate at temperatures below 4? K |
| + !regulate at temperatures below ~2 K |
| Line 61 was replaced by lines 63-64 |
| - * set the sorb heater loop to regulate the samlow sensor (channel B) |
| + * disable the sample heater |
| + {{ts set 0}} |
| At line 62 added 1 line. |
| + * set the sorb heater loop to regulate the samlow sensor (channel B) |
| Removed line 66 |
| - |
| At line 68 added 2 lines. |
| + * when oszillating, you might need to reduce the prop parameter from the default value 500 to e.g. 100 |
| + {{ts setsorb/prop 100}} |
| Removed line 70 |
| - !regulate a T > 4 K |
| Removed line 72 |
| - set the sorb heater loop to regulate the sorb (channel C) |
| At line 73 added 3 lines. |
| + !regulate in T range 2 K < T < 20 K |
| + |
| + set the sorb heater loop to regulate the sorb (channel C) |
| Lines 77-78 were replaced by line 83 |
| - |
| - {{tt setsorb 15}} |
| + {{ts setsorb 15}} |
| At line 83 added 5 lines. |
| + !higher temperatures and warmup (e.g. 80 K) |
| + |
| + {{tt set 80}} |
| + |
| + For a more precise control {{tt set}} might be set to value slightly lower the the desired T, and {{ts set}} to the desired target. |