Difference between version 16 and version 7:
Line 45 was replaced by line 45
- __sea_commands__
+ !!Sea/SICS Commands
Line 47 was replaced by line 47
- ''tt set <value>'' ...setpoint of loop 1 in K
+ !condense
Line 49 was replaced by lines 49-50
- ''tt set2 <value>'' ... setpoint of loop 2 in K
+ * disable the sample heater
+ {{ts set 0}}
At line 50 added 42 lines.
+ * set the sorb heater loop to regulate the sorb (channel C)
+ {{ts setsorb/channel C}}
+
+ * set the sorb heater to 30 K
+ {{ts setsorb 30}}
+
+ * wait 15 min, then disable sorb heating
+ {{ts setsorb 0}}
+
+ !regulate at temperatures below ~2 K
+
+ * disable the sample heater
+ {{ts set 0}}
+
+ * set the sorb heater loop to regulate the samlow sensor (channel B)
+ {{ts setsorb/channel B}}
+
+ * regulate at 0.5 K
+ {{ts setsorb 0.5}}
+
+ * when oszillating, you might need to reduce the prop parameter from the default value 500 to e.g. 100
+ {{ts setsorb/prop 100}}
+
+
+
+ !regulate in T range 2 K < T < 20 K
+
+ set the sorb heater loop to regulate the sorb (channel C)
+ {{ts setsorb/channel C}}
+
+ * regulate the sorb at an intermediate T (15 K)
+ {{ts setsorb 15}}
+
+ * regulate at 4.5 K
+
+ {{ts set 4.5}}
+
+ !higher temperatures and warmup (e.g. 80 K)
+
+ {{tt set 80}}
+
+ For a more precise control {{tt set}} might be set to value slightly lower the the desired T, and {{ts set}} to the desired target.