Line 45 was replaced by line 45 |
- !!Sea/SICS COmmands |
+ !!Sea/SICS Commands |
At line 48 added 3 lines. |
+ * disable the sample heater |
+ {{ts set 0}} |
+ |
Line 52 was replaced by line 55 |
- * set the sorb heater to 30 K and adjust the prop parameter for this |
+ * set the sorb heater to 30 K |
Removed lines 54-56 |
- {{ts setsorb/prop 500}} |
- |
- * wait 15 min then stop heating the sorb |
At line 57 added 1 line. |
+ * wait 15 min, then disable sorb heating |
Line 60 was replaced by line 61 |
- !regulate at temperatures below 4? K |
+ !regulate at temperatures below ~2 K |
Line 62 was replaced by lines 63-64 |
- * set the sorb heater loop to regulate the samlow sensor (channel B) |
+ * disable the sample heater |
+ {{ts set 0}} |
At line 63 added 1 line. |
+ * set the sorb heater loop to regulate the samlow sensor (channel B) |
Removed line 67 |
- |
Lines 70-71 were replaced by line 72 |
- * when oszillating, you might need to reduce the prop parameter |
- |
+ * when oszillating, you might need to reduce the prop parameter from the default value 500 to e.g. 100 |
Line 76 was replaced by line 77 |
- !regulate a T > 4 K |
+ !regulate in T range 2 K < T < 20 K |
Removed line 79 |
- |
Lines 83-84 were replaced by line 83 |
- |
- {{tt setsorb 15}} |
+ {{ts setsorb 15}} |
At line 89 added 5 lines. |
+ !higher temperatures and warmup (e.g. 80 K) |
+ |
+ {{tt set 80}} |
+ |
+ For a more precise control {{tt set}} might be set to value slightly lower the the desired T, and {{ts set}} to the desired target. |