Difference between version 11 and version 7:
Lines 13-15 were replaced by lines 13-27
- Comparison of the test at PSI (20 May 05) with test result at
- ILL (at 2 K, the time scale is different). For PSI, the same
- channel as at ILL is shown (RhFe Sensor)
+ Comparison of the test at PSI at 20 May 05 (left) with the test
+ result at ILL (right, different time scale). For PSI, the same
+ channel as at ILL is shown (RhFe Sensor).
+
+ __Conclusions from the Test__
+
+ * The stability is comparable to the results at the ILL (Feb 2003).
+ It is about +- 0.1 K below 4 K, an better than +- 0.05 K
+ above.
+ * Control on the cold valve is delicate, but possible. After cool down
+ it takes some time (2 hours) until a stable control on the cold
+ valve controller is possible.
+ * Further tests on TriCS with cradle moving are needed. It has to be
+ proven that the test results can be reproduced on the instrument.
+