- Test Results |
+ !Test Result February 2003 at ILL |
At line 2 added 4 lines. |
+ [Hef4cTestResults/he0.gif] |
+ |
+ !Test Results on TriCS Flow Cryostat from 20 May 2005 |
+ |
At line 6 added 16 lines. |
+ ---- |
+ [Hef4cTestResults/cmp.gif]\\ |
+ Comparison of the test at PSI at 20 May 05 (left) with the test |
+ result at ILL (right, different time scale). For PSI, the same |
+ channel as at ILL is shown (RhFe Sensor). |
+ |
+ __Conclusions from the Test__ |
+ |
+ * The stability is comparable to the results at the ILL (Feb 2003). |
+ It is about +- 0.1 K below 4 K, an better than +- 0.05 K |
+ above. |
+ * Control on the cold valve is delicate, but possible. After cool down |
+ it takes some time (2 hours) until a stable control on the cold |
+ valve controller is possible. |
+ * Further tests on TriCS with cradle moving are needed. It has to be |
+ proven that the test results can be reproduced on the instrument. |