Diff: HEOX

Search SINQ Wiki:
SINQ LIN

SINQ Wiki
- Main Page
- Search SINQ Wiki
- Sample Environment
- Probenumg. Intern
- Troubleshooting SICS

This Page
- Page Info
- Printer Friendly

Referenced by
DilutionCryostats

Wiki Info
- Unused pages
- Undefined pages
- RecentChanges
- Page Index
- System Info
- JSPWiki Docu
- SandBox
- OneMinuteWiki
- Create a New Page




JSPWiki v2.0.52


Difference between version 18 and version 14:
At line 29 added 1 line.
+ LS340 Settings:
At line 30 added 7 lines.
+ * Sensor A : 1K-pot
+ * Sensor B : Sorb. (thats the cryo pump for the 3He)
+ * Sensor C : 3He-pot Cernox for 1.2K<T<320K
+ * Sensor D : 3He-pot Speer for T<10K
+ * Sorb heater: 34 Ohm
+ * 3He-pot heater: 200 Ohm
+
Line 33 was replaced by line 41
- __Operation with two heaters:__
+ !Operation with two heaters:
At line 37 added 15 lines.
+ The switch on the heater cable allows to operate the Sorb on loop 1 (the main loop) and simultaneously
+ heat the 3He-pot with loop 2 and visa versa.
+
+ For low Temperature experiments <10K it is recommended to use: Sorb = loop1, 3He = loop2 while for higher
+ temperatures Sorb = loop2, 3He = loop1 should be used. ''note:'' for the high temperatur setting the Cernox (channel C)
+ should be used as the primary thermometer
+
+ You can operate the system by regulating the Sorb and the 3He-pot temperature independently (controlmode 0)
+ or by using both heaters cooperatively to control the 3He-pot temperature (controlmode 1) in this case the setpoints
+ for both loops should have the same value.
+
+ __INPORTANT:__ both heaters are disabled if any setpoint is exactly zero. In order to control the Temperature using only
+ 1 heater the other set value should be small (not reachable)
+
+
Line 40 was replaced by line 63
- ''tt controlmode <n>'' ...sets control channel of loop 1 to B (n=1) or D (n=0)
+ ''tt controlmode <n>'' ...sets control channel of loop 1 to B (n=0) or D (n=01)
Line 48 was replaced by line 71
- (e.g. tt send cset 2:D,1,1 sets loop 2 to control the He-3 pot heater
+ (e.g. tt send cset 2:D,1,1 sets loop 2 to control the 3He-pot heater

Back to HEOX, or to the Page History.