At line 29 added 1 line. |
+ LS340 Settings: |
At line 30 added 7 lines. |
+ * Sensor A : 1K-pot |
+ * Sensor B : Sorb. (thats the cryo pump for the 3He) |
+ * Sensor C : 3He-pot Cernox for 1.2K<T<320K |
+ * Sensor D : 3He-pot Speer for T<10K |
+ * Sorb heater: 34 Ohm |
+ * 3He-pot heater: 200 Ohm |
+ |
Line 33 was replaced by lines 41-42 |
- __Operation with two heaters:__ |
+ !Operation with two heaters: |
+ |
At line 36 added 15 lines. |
+ The switch on the heater cable allows to operate the Sorb on loop 1 (the main loop) and simultaneously |
+ heat the 3He-pot with loop 2 and visa versa. |
+ |
+ For low Temperature experiments <10K it is recommended to use: Sorb = loop1, 3He = loop2 while for higher |
+ temperatures Sorb = loop2, 3He = loop1 should be used. ''note:'' for the high temperatur setting the Cernox (channel C) |
+ should be used as the primary thermometer |
+ |
+ You can operate the system by regulating the Sorb and the 3He-pot temperature independently (controlmode 0) |
+ or by using both heaters cooperatively to control the 3He-pot temperature (controlmode 1) in this case the setpoints |
+ for both loops should have the same value. |
+ |
+ __INPORTANT:__ both heaters are disabled if any setpoint is exactly zero. In order to control the Temperature using only |
+ 1 heater the other set value should be small (not reachable) |
+ |
+ |
Lines 39-42 were replaced by lines 63-69 |
- ''tt controlmode'' <n> ...sets control channel of loop 1 to B (n=1) or D (n=0) |
- ''tt set2'' <n.n> ... setpoint of loop 2 in K |
- ''tt send pid'' n:P,I,D ... set PID values of loop n (1 or 2) |
- ''tt send cset'' n:<channel>,<unit>,<off/on>,<power enable> sets the channel (A,B,C,D) |
+ ''tt controlmode <n>'' ...sets control channel of loop 1 to B (n=0) or D (n=01) |
+ |
+ ''tt set2 <n.n>'' ... setpoint of loop 2 in K |
+ |
+ ''tt send pid n:P,I,D'' ... set PID values of loop n (1 or 2) |
+ |
+ ''tt send cset n:<channel>,<unit>,<off/on>,<power enable>'' ...sets the channel (A,B,C,D) |
Line 44 was replaced by lines 71-72 |
- (e.g. tt send cset 2:D,1,1 sets loop 2 to control the He-3 pot heater |
+ (e.g. tt send cset 2:D,1,1 sets loop 2 to control the 3He-pot heater |
+ |
At line 45 added 1 line. |
+ |