Diff: FurnaceList

Search SINQ Wiki:
SINQ LIN

SINQ Wiki
- Main Page
- Search SINQ Wiki
- Sample Environment
- Probenumg. Intern
- Troubleshooting SICS

This Page
- Page Info
- Printer Friendly

Referenced by
SampleEnvironment

Wiki Info
- Unused pages
- Undefined pages
- RecentChanges
- Page Index
- System Info
- JSPWiki Docu
- SandBox
- OneMinuteWiki
- Create a New Page




JSPWiki v2.0.52


Difference between version 19 and version 9:
Lines 1-4 were replaced by lines 1-4
- ||Name||Description ||T-range||sample\\space\\diameter||Thread on stick
- |[FS]|small furnace |350...700 K |25*|M6 female
- |[FT]|tantalum furnace|375...1400 K|50|M6 female
- |[FI]|ILL-type furnace (currently out of operation)|400...1800 K|40|M8 male
+ ||Name||Description ||T-range||sample\\space\\diameter||max.\\sample\\diameter||Thread on stick
+ |[FS]|small furnace |350...700 K |25*| |M6 female
+ |[FT]|tantalum furnace|375...1400 K|50|45|M6 female
+ |[FI]|ILL-type furnace|500...1800 K|40|35|M8 male
At line 22 added 1 line.
+ * __never use Cadmium or Indium in a furnace__, even not at relatively 'low' temperatures
Lines 24-25 were replaced by lines 25-35
- The picture below shows what might happen, if these rules are not
- obeyed. [FurnaceList/moltenSample.jpg]
+ The pictures below shows what might happen, if these rules are not
+ obeyed.
+
+ [FurnaceList/moltenSample.jpg]
+ \\Picture 1: Reaction of the sample with the container.
+
+ [FurnaceList/ftCd1.jpg]
+ [FurnaceList/ftCd2.jpg]
+ [FurnaceList/ftCd3.jpg]
+ [FurnaceList/ftCd4.jpg]
+ \\Picture 2,3,4,5: Molten and condensed Cadmium.

Back to FurnaceList, or to the Page History.