Diff: FT

Search SINQ Wiki:
SINQ LIN

SINQ Wiki
- Main Page
- Search SINQ Wiki
- Sample Environment
- Probenumg. Intern
- Troubleshooting SICS

This Page
- Page Info
- Printer Friendly

Referenced by
FurnaceList

Wiki Info
- Unused pages
- Undefined pages
- RecentChanges
- Page Index
- System Info
- JSPWiki Docu
- SandBox
- OneMinuteWiki
- Create a New Page




JSPWiki v2.0.52


Difference between version 16 and version 15:
Line 1 was replaced by line 1
- !Tantalum Furnace
+ !!Tantalum Furnace
Line 6 was replaced by line 6
- !Current Issues
+ !Heater power table
Lines 8-11 were replaced by lines 8-11
- The material of the heater element and the radiation shield will be exchanged
- by niobium. The reason is, that niobium seems to have some better properties,
- and that the furnace [FI] uses also niobium and therefore the purchase of
- materials will be easier. ''Anyhow, the furnace is operational''.
+ |T//[[K] |Heater Power//[[W]|U[[V]|Iprim//[[A]|I2nd//[A]|Prop|Int
+ |1400|2730|152|17.8|4 x 70.4|10|5
+ |1300|2060|130|15.8|4 x 57 |9|5
+ |1200|1500|110|14.1|4 x 46 |8|5
At line 12 added 2 lines.
+ Temperature of the connections at 1400 K: 90-100 deg Celsius
+

Back to FT, or to the Page History.