Difference between version 6 and version 1:
At line 2 added 2 lines.
+ * connect the analog output from the AVS with
+ the LakeShore 340
Line 5 was replaced by lines 7-10
-
+ * if the temperature is not shown, there might be a electronic
+ ground problem. Try to connect the LakeShores power cable to a
+ different phase. Warning: on channel C (sample), temperatures
+ beyond 20 K are not precise and are very noisy.
Removed line 9
-
Removed line 12
-
Removed line 14
-
At line 16 added 5 lines.
+ !Setup for temperatures beyond 1.5 K
+
+ * put a litte bit of exchange gas (1 mbar or less) into the IVC
+ * in ''tecs'' choose ''device variox/dil''
+ * set the temperature via ''tecs'' or SICS

Back to ControlTemperatureDIL, or to the Page History.