Setup for temperature control below 1.5 K

  • connect the analog output from the AVS with the LakeShore 340
  • on the AVS, choose all settings marked with a red dot
  • if the temperature is not shown, there might be a electronic ground problem. Try to connect the LakeShores power cable to a different phase. Warning: on channel C (sample), temperatures beyond 20 K are not precise and are very noisy.
  • on the IDR set the mixing chamber power to zero, switch the mixing chamber heater on and put the heater power range to 20mW
  • connect the heater cable from the IDR to the LakeShore 340 "Analog Out2" (coaxial cable)
  • in tecs choose device vardil
  • set the temperature via tecs or SICS

Setup for temperatures beyond 1.5 K

  • put a litte bit of exchange gas (1 mbar or less) into the IVC
  • in tecs choose device variox/dil
  • set the temperature via tecs or SICS