| At line 2 added 9 lines. |
| + It is highly recommended to use the automatic condensing scripts in SEA. |
| + Click on "sorbpump", when "readyforpump", make sure that the IVC is connected |
| + to a running turbo pump, and the valve and the IVC space is open. |
| + |
| + The automatic script is doing a sligthly different condensing procedure than |
| + described below. |
| + |
| + * If not yet done, the IVC has to be pumped. Do not do that before the sorb T |
| + is at 4K. |
| Line 6 was replaced by line 15 |
| - * Cool the 1 K pot to a temperature below 3 K. With the |
| + * Cool the VTI ("1 K pot" thermometer) to a temperature below 3 K. With the |
| Lines 8-13 were replaced by lines 17-19 |
| - It might be necessary to open the needle valve more than |
| - usual in order to fill the pumping volume with some |
| - amount of liquid helium. Be aware, if the amount of |
| - liquid is too high, it might take hours to get a stable |
| - flow. |
| - * Open the 3 valves in the pumping line and the 2 valves in |
| + Open the needle valve to get a pressure at the 65 m3 |
| + pump of about 12 mbar. |
| + * Open the valves in the pumping line and in |
| Line 15 was replaced by line 21 |
| - * Open the green valve marked "out" below the dump. The dump is the |
| + * Open the green valves marked "out" and "in" on the dump. The dump is the |
| Lines 18-19 were replaced by lines 24-27 |
| - RAISE). |
| - * Open valve 9 (while pressing on the valve button, press |
| + RAISE). |
| + * Make sure that the heater on the mixing chamber is off and that |
| + the setpoint is zero. Switch off the still heater also. |
| + * Open valve 9 (while pressing on the valve button, press |
| Lines 21-22 were replaced by lines 29-30 |
| - * The reading on G2 is proportional to the amout of mixture in the |
| - dump. It should read something around 660 mbar. |
| + * The reading on G2 is proportional to the amount of mixture in the |
| + dump. It should read something around 660(DIL1) or 400(DIL2) mbar. |
| Lines 26-29 were replaced by lines 34-36 |
| - * Open a little valve 12A (4.0 % to start). |
| - * While the pressure reading on G1 should always be below about 400 mbar, |
| - Valve 12A can bae opened gradually, and opened completely, when the |
| - reading on G2 drops below 350 mbar. |
| + * Open V12A a little (4 %) to allow the mixture pass the trap slowly. |
| + * When the pressure reading on G1 stabilizes or decreases, open valve 12A |
| + completely. |
| Lines 34-40 were replaced by lines 41-48 |
| - |
| - * After some time (about 30 min when using the Variox) the pressure on |
| - G1 falls below 150 mbar, and it is time to pump out the dump: close |
| - valve 9, and open valve 14 for a short time (close it when P1 raises |
| - above 50 mbar). As soon as the pressure on P1 drops down, you might open |
| - and close valve 14 again, but make sure that the pressure on G1 does |
| - not increase above 350 mbar. |
| + * After some time (depends on the temperature of the mix. chamber) |
| + the mixing chamber should reach about 1 K, and then |
| + the pressure on G1 stabilizes at a low value. |
| + Now you may start to pump out the dump: close valve 9, |
| + and open valve 14 for a short time |
| + (close it when P1 raises above 50 mbar). |
| + As soon as the pressure on P1 drops down, you might open |
| + and close valve 14 again. Do that several times. |
| Line 42 was replaced by line 50 |
| - raise above 50 mbar. Leave the valve open until the P1 drops below 0.1. |
| + raise above 50 mbar. Leave the valve open until P1 drops below 1. |
| Removed line 48 |
| - |
| Lines 50-51 were replaced by lines 57-58 |
| - the flow on the condenser line might falls below 100 (black display below |
| - the valve panel). |
| + the flow on the condenser line might fall below 100 (black display below |
| + the valve panel, on DIL1 only). |
| At line 53 added 2 lines. |
| + |
| + [DIL1/igh.png] |